Benzene / Toluene/ Ethanol in

Gasoline

ASTM Method D3606

Present solution is problematic:

Big issues with separations and

peak shapes

Modified ASTM D3606 Method using the

D3606 Column Set

• First part of eluting peaks is sent to the second, highly selective

column, where the alcohols and aromatics are separated

• Higher boiling compounds are back-flushed

• Unique column set, only from Restek

Uses a SET of packed columns:

Material

Siltek deactivated SS

ID / OD

2mm / 1/8”

Lengths

6’ and 16’ long

Phases

Rtx-1 and proprietary

New D3606 Method

Columns :Rt™-D3606ColumnSet (cat.#83606)

column1 : Rt™-D3606-1:6'x1/8"OD (1.8m x2mm ID), nonpolar Rtx®-1

column2 : Rt™-D3606-2:16'x1/8"OD (4.9m x 2mm ID), proprietarypacking

Injection : 0.05µL,direct injection; Inj. temp. :200°C

Car. Gas : He, constant flow;Flow

rate:20mL/min.

Oven

:135°C,

Det. : FID, temp.:250°C

Highest possible

resolution!!

Unique Restek solution