Restek 2011/12 Chromatography Catalog - page 669

669
GC CHROMATOGRAMS |
CLINICAL, FORENSIC & TOXICOLOGY
GHB/GBL
γ
-butyrolactone & 1,4-butanediol
Rtx®-BAC1 & Rtx®-BAC2
Columns:
Rtx
®
-BAC1, 30m, 0.32mm ID, 1.8µm (cat.# 18003)
Rtx
®
-BAC2, 30m, 0.32mm ID, 1.2µm (cat.# 18002)
Sample:
100µg/mL each in methanol
Inj.:
1.0µL split (split ratio 1:10), 4mm single gooseneck
inlet liner with wool (cat.# 22405)
Inj. temp.:
240°C
Carrier gas:
helium, constant flow
Flow rate:
3.0mL/min.
Oven temp.:
50°C to 240°C @ 10°C/min. (hold 2 min.)
Det.:
FID @ 240°C
1. 2,3-butanediol
2. 1,2-butanediol
3. 1,3-butanediol
4. 1,4-butanediol
5.
γ
-butyrolactone
GC_PH00808
GC_PH00809
Rtx®-BAC1
Rtx®-BAC2
γ
-hydroxybutyrate (GHB) and
γ
-butyrolactone (GBL)
Rtx®-BAC1 and Rtx®-BAC2 (dual column analysis)
GC_PH00870
GC_PH00871
Columns:
Rtx
®
-BAC1, 30m, 0.32mm ID, 1.8µm (cat.# 18003) and
Rtx
®
-BAC2, 30m, 0.32mm ID, 1.2µm (cat.# 18002),
connected via universal “Y” Press-Tight
®
connector (cat.# 20405)
Sample:
GHB, GBL,
α
-methylene-
γ
-butyrolactone (AMGBL), 10µg/mL each in water
Inj.:
1.0mL headspace, split (split ratio 1:10), 1mm split inlet liner (cat.# 20972)
Inj. temp.:
200°C
Carrier gas:
helium, constant pressure
Linear velocity:
44cm/sec. @ 50°C
Oven temp.:
50°C (3 min.) to 150°C @ 20°C/min. (hold 7 min.)
Det:
FID @ 240°C
Headspace autosampler: Teledyne/Tekmar HT3
Sample/platen temp.:
100°C
Sample equilibration:
15 min.
Mixing time:
5 min.
Vial pressure:
10psig
Vial pressurization
time:
2 min.
Loop fill time:
2 min.
Transfer line temp.:
120°C
1.
γ
-hydroxybutyrate (GHB)/
γ
-butyrolactone (GBL)
2.
α
-methylene-
γ
-butyrolactone (AMGBL)
Rtx®-BAC1
Rtx®-BAC2
Sharp peaks and baseline
resolution in less than 5 minutes—
12 samples per hour!