MRC-2013-Catalog - page 620

620
S
SPECTROPHOTOMETERS
SPECTRO-UV6 Series is an advanced double beam design consisting of six
models.
Stand-alone model:
SPECTRO-UV61 with 1.8nm fixed bandwidth
SPECTRO-UV63 with 1.0 fixed bandwidth.
SPECTRO-UV61S with variable bandwidth: 0.5/1/2/5nm
SPECTRO-UV6 Series, Double Beam Spectrophotometer
Model
SPECTRO-UV61
SPECTRO-UV61PC
SPECTRO-UV63
SPECTRO-UV63PC
SPECTRO-UV61S
SPECTRO-UV61PCS
Wavelength range
190-1100nm
Spectral bandwidth
1.8NM
1nm 0.5/1/1/4nm
Optical system
Double beam, grating 1200 lines/mm
Wavelength accuracy
±0.3nm
Wavelength repeatability
0.2nm
Scan Speed
Hi, MED., LOW., MAX.3000nm/min
Photometric accuracy
±0.3%T or ±0.002A@1A
Photometric range
0-200%T, -0.3 -3A
Stray light
0.05%T@220nm, 360nm
Stability
±0.001A/h @500nm
Display
5 inches LCD(320*240 dots)
Baseline Flatness
±0.001A
Standard cell holder
Standard 10mm single cell holder(2 pcs)
Sample Compartment
Standard 10mm pathlength cuvette
Light source
tungsten & Deuterium lamp(pre-aligned)
Output
USB Port & parallel port (printer)
Power requirement
AC 110/220V 50/60Hz
Dimensions (WxDxH)
600x450x200mm
Weight
22kg
Fixed or variable slits (bandwidths)
For Stand-alone models, all software methods are included as built-in
standard; this eliminates the need of software.
Online software upgrade via internet helps to keep it updated.
Data Download-to-PC software expands the data storage to unlimited.
The stand-alone models has 5 inch screen and the PC models has
UVNis Analyst software. Stand-alone models of SPECTRO-UV6 Series
have the same functions as SPECTRO-UV3 Series, see next page for details.
SPECTRO-UV6 Series
Other specifications of the six models are
almost the same except bandwidth.
The two detectors measure sample and
reference respectively and
simultaneously for optimizing
measurement accuracy, They provide
excellent performance for measurements
in the range of 190 to 1100nm, They are
suitable for pharmaceutical, biochemical
and clinical lab applications as well as
routine applications such as
quantitative analyses, kinetics,
wavelength scan, multiple components
and DNA/Protein, PC Windows
application software make these
instruments versatile.
All instruments provide excellent
performance for measurements.
All methods are included as built-in standard; this eliminates the
need of software. Online software update via Internet.
The local control software includes functions such as: Photometry,
Quantitative, Wavelength Scan, Kinetics, DNA/Protein,
Multi-wavelength and System Utilities.
Standard Curve·
Up to 10 standard solutions may be used to establish calibration equation
curve. There is a choice of four methods for fitting curve through the
calibration points: Linear fit, Linear fit through zero, square fit and cubic fit.
Wavelength Scan
The Wavelength Scan intervals are 0.1, 0.2, 0.5, 1, 2, 5nm, and High, Medium
and Low scan speeds are available. Scan speeds vary from 100 to 1000
nm/min. Wavelengths are scanned from high to low so that the instrument
stand-by at high wavelength. This minimizes the degradation of UV sensitive
samples. Precise control of filter and lamp changes means that their effects
are not seen on the final scan. Post-run manipulation includes re-scaling
axes, curve tracking and peak picking.
SPECTRO-UV6 Series Local Control Software
VIS - UV
1...,610,611,612,613,614,615,616,617,618,619 621,622,623,624,625,626,627,628,629,630,...738
Powered by FlippingBook