SPECTROPHOTOMETERS
SPECIFICATIONS
Main Specification
Wavelength range
190-900nm
Wavelength accuracy
Better than ±0.25nm
Resolution
Two spectral lines of Mn at 279.5nm and 279.8nm can be
separated with the spectral bandwidth of 0.2nm and
valley-peak energy ratio less than 30%.
Baseline stability
0.004A/30min
Background correction
The D2 lamp background correction capability at 1A is
better than 30 times.
The S-H background correction capability at 1.8A is better
than 30 times.
Light Source System
Lamp turret
Motorized 6-lamp turret (Two high performance HCLs can be
mounted on the turret to increase the sensitivity in flame
analysis.)
Lamp current adjustment
Wide pulse current: 0~25mA,
Narrow pulse current: 0~10mA.
Lamp power supply mode
400Hz square wave pulse; 100Hz narrow square wave pulse +
400Hz wide square wave pulse.
Optical System
Monochromator
Single beam, Czerny-Turner design grating monochromator
Grating
1800 l/mm
Focal length
277mm
Blazed wavelength
250nm
Spectral bandwidth
0.1 nm, 0.2nm, 0.4nm, 1.2nm, auto switch over
Flame Atomizer
Burner
10cm single slot all-titanium burner
Spray chamber
Corrosion resistant all-plastic spray chamber.
Nebulizer
High efficiency glass nebulizer with metal sleeve.
sucking up rate: 6-7mL/min
Emission burner provided
Graphite Furnace
Temperature range
Room temperature-3000°C
Heating rate
2000°C/s
Graphite tube dimensions
28mm (L) x 8mm (OD)
Characteristic mass
Cd
≤
0.8*10
-12
g, Cu
≤
5*10
-12
g, Mo
≤
1*10
-11
g
Precision
Cd
≤
3%, Cu
≤
3%, Mo
≤
4%
Detection and Data
Processing System
Detector
R928 photomultiplier with high sensitivity and wide
spectral range.
Software
Under Windows operating system
Analytical method
Working curve auto-fitting; standard addition method;
automatic sensitivity correction;
automatic calculation of concentration and content.
Repeat times
1~99 times. automatic calculation of mean value, standard
deviation and relative standard deviation
Multi-task function
Sequential determination of multi-elements in the same sample.
Condition reading
With model function
Result printing
Measurement data and final analytical report printout, editing
with Excel.
Standard RS-232 serial port communication
Graphite Furnace
Autosampler
Sample tray capacity
55 sample vessels and 5 reagent vessels
Vessel material
Polypropylene
Vessel volume
3ml for sample vessel. 20ml for reagent vessel
Minimum sampling volume
1µl
Repeatable sampling times
1~99 times
Sampling system
Accurate dual pump system. with 100µI and 1 ml injectors.
Characteristic
Concentration and
Detection Limit
Air-C 2 H 2 flame
Rich oxygen Air-C 2 H 2 flame
Cu: Characteristic concentration
≤
0.025 mg/L,
Detection limit
≤
0.006mg/L;
Ba: Characteristic concentration
≤
0.22mg/L
AI: Characteristic concentration
≤
0.4mg/L
Function Expansion
Hydridee vapor generator can be connected for Hydridee analysis.
Dimensions and Weight
Main unit
107X49x58cm,140kg
Graphite furnace
42X42X46cm, 65kg
Autosampler
40X29X29cm, 15kg
Atomic Absorption