restektn07 - page 331

Figure1
CarbamatepesticidesonanUltraCarbamatecolumn.
Peak List:
1. aldicarb sulfone
2. aldicarb sulfoxide
3. oxamyl
4. methomyl
5. 3-hydroxycarbofuran
6. aldicarb
7. propoxur
8. carbofuran
9. carbaryl
10. methiocarb
11. 4-bromo-3,5-dimethylcarbamate
Column:
Ultra Carbamate
Cat.#:
9177355
Dimensions:
50 x 4.6mm
Particle size:
3µm
Pore size:
100Å
Conditions:
Mobile phase:
A: 90:10water:methanol
B: 90:10methanol:
acetonitrile
Time (min.) %B
0
10
10
90
Flow:
1.5mL/min.
Temp.:
27°C
Det.:
UV@ 220nm
LC_0225
Sample:
Inj.:
5µL cat.# 32274 and
cat.# 32273mixed 50:50
Conc.:
50µg/mL each
Solvent:
methanol
CarbamatesbyHPLC
UltraCarbamateColumns
PhysicalCharacteristics:
*For post-column derivatization / fluorescence detection appli-
cations for a 4.6mm ID column the total system dead volume,
including the post-column reactor, must be less than 650µL. For
standard post-column reactor systems, we recommend a 250 x
4.6mm, 5µm column. Call Restek technical service formore
information.
3µmColumn
cat.#
price
50mm (2.1mm ID)
9177352
$438
50mm (4.6mm ID)
9177355
$454
5µmColumn
cat.#
price
250mm (4.6mm ID)
9177575
$438
ordering note
For guard cartridges for these columns, visit ourwebsite at
particle size: 3µmor
5µm, spherical
pore size: 100Å
pH range: 2.5 to7.5
temperature limit: 80°C
aldicarb
aldicarb sulfone
aldicarb sulfoxide
carbaryl (Sevin
®
)
carbofuran
3-hydroxycarbofuran
methiocarb
methomyl
oxamyl
propoxur (Baygon
®
)
531.1CarbamatePesticideCalibrationMixture
(10 components)
100µg/mL each inmethanol, 1mL/ampul
cat.# 32273 (ea.)
$41
aldicarb
aldicarb sulfone
aldicarb sulfoxide
carbaryl (Sevin
®
)
carbofuran
3-hydroxycarbofuran
methiocarb
methomyl
1-naphthol
oxamyl
propoxur (Baygon
®
)
531.2CarbamatePesticideCalibrationMixture
(11 components)
100µg/mL in acetonitrile, 1mL/ampul
cat.# 32435 (ea.)
$46
4-bromo-3,5-dimethylphenyl-N-methylcarbamate (BDMC)
InternalStandard
100µg/mL inmethanol, 1mL/ampul
cat.# 32274 (ea.)
$23
aldicarb sulfoxide 100µg/mL
BDMC
10
3-hydroxycarbofuran
2
methiocarb
20
531.1PerformanceCheckMix
Inmethanol, 1mL/ampul
cat.# 32275 (ea.)
$41
Carbamates are widely used insecticides that pose a health risk as endocrine
disruptors.OurUltraCarbamate column, in a 50mm length, separates com-
mon carbamates in less than10minutes (Figure 1), significantly less than the
time required by traditional C18 columns. In addition to the best column
choice for the analysis,weoffer referencemixes forMethod531 carbamates, a
performance check mix, and the specified internal standard, 4-bromo-3,5-
dimethylphenyl-N-methylcarbamate (BDMC).
RecommendedColumns
AnalyticalReferenceMaterials
11
free
literature
Simple, SensitiveHPLC/UVAnalysis for Paraquat andDiquat
These highly charged quaternary amines are poorly retained on alkyl stationary phases. Using only
acetonitrile, water, and a solvation-blocking reagent, our separation system alters the interactions
among analyte, mobile phase, and stationary phase, and promotes solubility of the analytes in the
stationary phase. In our system, the detection limit is 6ppb for either herbicide, and the analysis is
completed in less than 10minutes. An optimized solid phase extraction cartridge concentrates the
herbicides for the analysis.
Lit. cat.# 580006