ASTM Method D-6352 is a gas chromatography
(GC) method developed for the determination of
petroleum distillates with a boiling point range of
174°C to 700°C. Often referred to as “high-temper-
ature simulated distillation,” this method requires a
capillary column capable of withstanding GC oven
by Neil Mosesman, GC Columns Product Marketing Manager
Fast, High-Temperature
Sim Dist Analysis
MXT
®
-1HT Sim Dist Capillary GC
Column and the GC Racer*
✔
Analysis time reduced by 75%, increases throughput.
✔
Meets all criteria of ASTM Method D-6352.
✔
Excellent peak shape for high-boiling compounds.
temperatures up to 430°C. This presents many
challenges for analysts because most capillary
columns are manufactured using polyimide-coated
fused silica tubing. At temperatures above 380°C,
even the best polyimide coating becomes brittle,
which leads to very short column lifetimes. In addi-
tion, the methyl silicone stationary phase recom-
mended in the method also must survive these high
temperatures.
The MXT
®
-1 HT Sim Dist column is a major
improvement in column technology for high-tem-
perature simulated distillation. By combining a new,
proprietary polymer synthesis technology, Siltek
™
deactivation, and rugged tubing, we developed a
Figure 1
The MXT
®
-1HT Sim Dist column exhibits
lower bleed and improved peak shape
compared to other columns used for
simulated distillation.
capillary column that meets all the criteria of ASTM
Method D-6352. This MXT
®
-1HT Sim Dist column is
available as a 5m, 0.53mm ID, 0.1µm film to con-
form to the requirements of this method. It exceeds
the resolution, peak shape, and bleed criteria for
hydrocarbons ranging up to C-110. Because the
MXT
®
-1HT Sim Dist column is coated with a 100%
dimethyl polysiloxane polymer, it will give the cor-
rect retention time/boiling point curve. The MXT
®
-
1HT Sim Dist column exhibits low bleed and excel-
lent inertness, and the rugged tubing will hold up to
temperatures in excess of 430°C.
To demonstrate the lower bleed and improved peak
shape of this innovative column, a Polywax
®
1000
reference material was analyzed using an MXT
®
-
1HT Sim Dist column and two other columns that
are commonly used for this application (Figure 1).
The MXT
®
-1HT Sim Dist column exhibits lower
bleed and improved peak symmetry compared to
other columns on the market.
As part of cost reduction efforts, many laboratories
try to reduce individual sample analysis times in the
interest of increasing overall throughput. High-tem-
perature simulated distillation analyses can take as
long as an hour, especially when samples contain
hydrocarbons up to C110. An effective technique to
reduce analysis time is to use rapid temperature
programming. Unfortunately, most GC systems have
temperature-programming limitations of 20°C to
25°C/min. To overcome these limitations, Restek
offers the GC Racer, an attachment to your Agilent
GC_PC00542
GC_PC00541
Sample:
0.2µL Polywax
®
1000 standard
(cat.# 36227);
Solvent:
carbon disulfide;
Injector:
on-column (track oven);
Carier gas:
heli-
um (constant pressure);
Head pressure:
1.0psi;
Linear velocity:
60cm/sec.;
Flow rate:
7.8mL/min.;
Detector:
FID @ 430°C;
Make-up gas flow:
40cc/min.;
Temp. program:
40°C to 430°C @
10°C/min. (hold 30 min.)
Brand Y
Brand X
Figure 2
Significantly reduce simulated distillation analysis time using the MXT
®
-1HT Sim Dist column
and the GC Racer attachment.
C44
C50
C60
C70
C80
C90
C100
C110
C120
C130
2
4
6
8
10
12
14
16
18
5m, 0.53mm ID, 0.10µm MXT
®
-1HT Sim Dist (cat.# 70100)
Sample:
Polywax 1000 (cat.# 36227);
Solvent:
carbon disulfide;
Sample size:
0.2µL;
GC:
HP5890 GC w/ GC Racer
™
system;
Injector:
on-column (track oven);
Carrier gas:
hydrogen;
Constant pressure:
1.0psi;
Detector:
FID @ 430°C;
Temp. program:
40°C to 430° C @ 60°C/min.
D6352-98 Polywax
®
Standards
These high molecular weight hydrocarbon waxes are useful for simulated distillation and other
high-temperature GC work.
Description
qty.
cat.#
Ea.
Polywax 500
1 gram
36224
Polywax 655
1 gram
36225
Polywax 850
1 gram
36226
Polywax 1000
1 gram
36227
0
10
20
30
40
50 min.
0
10
20
30
40
50 min.
0
10
20
30
40
50 min.
78pA
85pA
GC_PC00543
GC_PC00530
C30
C110
C30
C110
MXT
®
-1HT Sim Dist
C30
C110
Wow!
28pA