Table of Contents Table of Contents
Previous Page  133 / 298 Next Page
Information
Show Menu
Previous Page 133 / 298 Next Page
Page Background

• 7 •

www.restekcorp.com

800-356-1688

Why use a guard column?

Capillary gas chromatography (GC) guard columns

protect analytical columns by trapping nonvolatile

residues, preventing them from collecting at the

front of the analytical column. These nonvolatile

residues may be very high molecular weight organic

compounds, inorganic salts, or particulates. If these

contaminants enter the analytical column, they can

cause adsorption of active compounds, retention

time drift, loss of resolution, and poor peak sym-

metry. When this contamination begins to affect

sample analysis, a small section of the analytical

column must be removed to restore proper per-

formance. Each time a section of the analytical col-

umn is removed, retention times change, and some

resolution is lost. By using a guard column and

removing contaminated loops from it instead of the

analytical column, the separation power of the ana-

lytical column remains intact.

When should a guard column be replaced?

The guard column should be replaced as it

becomes contaminated with nonvolatile residue. At

this point, the performance of the entire chromato-

graphic system will begin to deteriorate. This nor-

mally is exhibited as a drastic decrease in the

response of active compounds and peak tailing.

What is the life expectancy of a guard

column?

The life expectancy of a guard column depends on

its length, the amount of nonvolatile residue in the

samples, and the nature of samples injected on the

column. When analyzing dirty samples, the guard

column becomes contaminated quickly. Normally,

contamination deposits in the first meter of the

guard column. If a short guard column (1m) is

used, it must be completely replaced when it

becomes contaminated. If a longer guard column

(5m) is used, the contaminated sections can be

removed without having to reconnect it to the ana-

lytical column.

What length guard column do I need?

A guard column should be long enough to keep

non-volatile residue from entering the column, but

short enough so that the analysis time is not dra-

matically increased. Five-meter guard columns are

more cost effective, reduce the frustrations of mak-

ing constant connections between shorter guard

columns and the analytical column. Ten-meter

guard columns often are used when analyzing very

dirty environmental samples. If a very long guard

column (>10 meters) is used, the residence time

of sample components increases, resulting in a

slightly longer analysis time. Guard columns over

30 meters long can cause peak distortion and a

loss in efficiency; they are not recommended. In

any case, it is important to adjust the column flow

rate to account for the length of the guard column

that is used, even though the guard column does

not have retention.

Selection Tips

Several types of guard columns are available. The choice of guard column should be made depending upon

the compatibility of the guard column type with the compounds of interest, or if the tubing is being used as a

transfer line to carry the sample from an inlet device to the column or from the column outlet to the detec-

tor. Choosing a guard column that is compatible with your compounds of interest ensures focused sample

bands, and good peak shape with minimal peak tailing.

Integra-Guard

Guard Columns

• Continuous length of tubing containing both the

guard column and the analytical column—no

connector required.

• Available in many phases.

• Guaranteed leak-free.

Specialized Deactivations

Siltek

Guard Columns

Revolutionary deactivation lowers endrin break-

down to less than 1%.

Inertness retained over a wide range of sample pH.

• Minimal bleed.

• Ideal for chlorinated pesticide analysis.

• Individually tested for chlorinated pesticide

breakdown.

Recommended for difficult matrix and reactive

compound analysis

Ideal for use as transfer lines.

• Recommended for use with Rtx

®

-CLPesticides,

Stx

®

-CLPesticides, Stx

®

-1HT, and Rtx

®

-TNT

columns.

Base-Deactivated Guard Columns

• Provides excellent inertness for the analysis of

basic compounds.

• Tested with basic amine test mix (chromatogram

included).

• Recommended for use with Rtx

®

-5 Amine,

Rtx

®

-35 Amine, and Stabilwax

®

-DB columns

General-Purpose Deactivations

In most cases, the standard IP tubing should be

chosen. The IP surface contains methyl, as well as

phenyl groups, making this surface compatible with

most common solvents.

Intermediate Polarity (IP) Tubing

• The most universal guard column tubing material.

• Phenylmethyl-deactivated surface provides

optimum compatibility for both polar and

non-polar compounds.

Polar-Deactivated Tubing

• Provides optimum wettability for polar com-

pounds.

• Minimizes peak splitting when using polar sol-

vents such as methanol and water.

• Uses a polyethylene glycol deactivation layer.

• Compatible with Stabilwax

®

, Rtx

®

-225, and

Rtx

®

-2330 capillary columns.

If methanol or water is the primary solvent, then

polar surfaces should be used such as our polar-

deactivated tubing. The polar-deactivated surface is

not resistant to harsh water vaporization, which

occurs when water in the liquid state is injected

onto the tubing surface and rapidly vaporized.

Hydroguard

Tubing

• Provides excellent inertness for water-based

samples.

• Reduces effects of dirty samples on column

performance.

• Reduces downtime and maintenance.

Hydroguard

tubing is preferred for situations

where there is harsh water vaporization. By using a

unique deactivation chemistry, the resulting high-

density surface is not readily attacked after an

aggressive hydrolysis treatment. The high-density

surface coverage effectively prevents water vapor

from reaching the fused silica surface beneath the

Hydroguard

deactivation layer.

GC Capillary Guard

Column Options

Save money by prolonging the lifetime of your analytical column.

Ensure reproducible analyses by keeping nonvolatile residue from

collecting at the front of the analytical column.

May improve sample focusing and separation efficiency.

by Christine Vargo, US Sales and Distribution Manager