Page Background

122

www.restek.com

GC COLUMNS |

PLOT COLUMNS

Fused Silica Alumina BOND Columns

Natural Gas on Rt®-Alumina BOND/KCl

GC_PC01091

Column

Rt®-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10 µm (cat.# 19760)

Sample

Natural gas

Injection

Inj. Vol.:

500 µL split

Liner:

2.0 mm ID single taper (cat.# 20795)

Inj. Temp.:

200 °C

Split Vent

Flow Rate:

50 mL/min

Oven

Oven Temp.:

45 °C (hold 1 min) to 200 °C at 10 °C/min (hold 8.5 min)

Carrier Gas

H

2

, constant pressure (8.0 psi, 55.2 kPa)

Linear Velocity:

45 cm/sec @ 45 °C

Detector

FID@ 200 °C

Make-up Gas

Type:

N

2

Data Rate:

20 Hz

Instrument

HP5890 GC

Peaks

1. Methane

2. Ethane

3. Propane

4. Isobutane

5.

n

-Butane

6. neo-Pentane

7. iso-Pentane

8.

n

-Pentane

9. Hexane isomers

10. Heptane isomers

11. Benzene

12. Octane isomers

Rt®-Alumina BOND/KCl Columns

(fused silica PLOT)

(KCl deactivation)

• Restek’s lowest polarity alumina column.

• Low moisture sensitivity reduces the need for frequent regeneration.

• Acetylene elutes before

n

-butane.

• Methyl acetylene (impurity in 1,3-butadiene) elutes before 1,3-butadiene.

• Stable to 200 °C.

similar

phases

GC-Alumina KCl, HP-PLOT Al

2

O

3

/KCl,

CP-Al

2

O

3

/KCl, Alumina-Chloride

Fused Silica Capillary & PLOT

Column Ferrule Guide

GC Column ID

Ferrule ID

0.15 mm

0.4

0.18 mm

0.4

0.25 mm

0.4

0.32 mm

0.5

0.53 mm

0.8

30-Meter

50-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.25 mm 4 µm to 200 °C

19776

0.32 mm 5 µm to 200 °C

19761

19762

0.53 mm 10 µm to 200 °C

19759

19760

We pack Restek quality into each

and every column we ship.