Page Background

110

www.restek.com

GC COLUMNS |

METAL MXT® CAPILLARY COLUMNS

General-Purpose Columns

MXT®-200 Columns

(Siltek®-treated stainless steel)

(midpolarity phase; Crossbond® trifluoropropylmethyl polysiloxane)

• General-purpose columns for solvents, Freon® fluorocarbons, alcohols, ketones,

silanes, and glycols. Excellent confirmation column with an Rtx®-5 column, for

phenols, nitrosamines, organochlorine pesticides, chlorinated hydrocarbons, and

chlorophenoxy herbicides.

• Temperature range: -20 °C to 400 °C.

• Equivalent to USP G6 phase.

• 4.5

"

standard coil diameter.

*Maximum temperatures listed are for shorter length columns. Longer columns may have a different maximum temperature.

MXT®-200 Structure

CH

3

CF

3

(CH

2

)

2

Si

O

MXT®-WAX Structure

H

H

C C O

H

H

Similar to: (trifluoropropyl)-methylpolysiloxane

also

available

Metal MXT®

PLOT Columns

See

page 129

.

15-Meter

30-Meter

60-Meter

ID

df

temp. limits*

cat.#

price

cat.#

price

cat.#

price

0.25 mm 0.50 µm -20 to 400 °C

75038

1.00 µm -20 to 310/360 °C

75053

0.53 mm 1.00 µm -20 to 290/360 °C 75052

75055

75058

1.50 µm -20 to 280/360 °C

75070

75073

3.00 µm -20 to 260/360 °C 75082

75085

75088

MXT®-WAX Columns

(Siltek®-treated stainless steel)

(polar phase; Crossbond® Carbowax® polyethylene glycol—provides oxidation resistance)

• General-purpose columns for FAMEs, flavor compounds, essential oils, amines,

solvents, xylene isomers, and U.S. EPA Method 603 (acrolein/acrylonitrile).

• Temperature range: 40 °C to 260 °C.

• Equivalent to USP G14, G15, G16, G20, and G39 phases.

• 4.5

"

standard coil diameter.

similar phases

DB-PSWAX, UAC-CW

*Maximum temperatures listed are for shorter length columns. Longer columns may have a different maximum temperature.

15-Meter

30-Meter

60-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

cat.#

price

0.25 mm 0.25 µm 40 to 250/260 °C 70620

70623

0.50 µm 40 to 260 °C

70638

0.28 mm 0.25 µm 40 to 250/260 °C

70624

0.50 µm 40 to 250/260 °C

70639

70642

1.00 µm 40 to 240/250 °C 70651

70654

70657

0.53 mm 0.25 µm 40 to 250/260 °C 70622

70625

0.50 µm 40 to 250/260 °C 70637

70640

1.00 µm 40 to 240/250 °C 70652

70655

70658

1.50 µm 40 to 230/250 °C 70666

70669

70672

2.00 µm 40 to 220/250 °C 70667

70670

Aromatics (Benzene/Toluene/Xylene) on MXT®-WAX

GC_PC00188

Column

MXT®-WAX, 30 m, 0.28 mm ID, 1.00 µm

(cat.# 70654)

Sample

BTEX Standard (cat.# 30051)

Injection

Inj. Vol.:

1.0 µL split (split ratio 100:1)

Inj. Temp.:

260 °C

Oven

Oven Temp.:

60 °C

Carrier Gas

H

2

, constant flow

Linear Velocity:

38 cm/sec

Detector

FID@ 260 °C

Notes

FID sensitivity: 3.2 x 10

-11

AFS

Peaks

1. Benzene

2. Toluene

3. Ethylbenzene

4.

p

-Xylene

5.

m

-Xylene

6.

o

-Xylene