Table of Contents Table of Contents
Previous Page  110 / 628 Next Page
Information
Show Menu
Previous Page 110 / 628 Next Page
Page Background

108

www.restek.com

GC COLUMNS |

METAL MXT® CAPILLARY COLUMNS

General-Purpose Columns

MXT®-5 Structure

CH

3

CH

3

Si

95%

O

Si

5%

O

Similar to: (5%-phenyl)-methylpolysiloxane

also

available

Metal MXT®

PLOT Columns

See

page 129

.

MXT®-5 Columns

(Siltek®-treated stainless steel)

(low-polarity phase; Crossbond® diphenyl dimethyl polysiloxane)

• General-purpose columns for drugs, solvent impurities, pesticides, hydrocarbons,

PCB congeners (e.g., Aroclor mixes), essential oils, and semivolatiles.

• Temperature range: -60 °C to 430 °C.

• Equivalent to USP G27, G36 phases.

• 4.5

"

standard coil diameter.

The diphenyl dimethyl polysiloxane stationary phase is the most popular GC station-

ary phase and is used in a wide variety of applications. All residual catalysts and low

molecular weight fragments are removed from the MXT®-5 polymer, providing a tight

monomodal distribution and extremely low bleed.

similar phases

DB-PS5, VF-5ht UltiMetal, UAC-5, UAC-5MS

*Maximum temperatures listed are for shorter length columns. Longer columns may have a different maximum temperature.

15-Meter

30-Meter

60-Meter

ID

df

temp. limits*

cat.#

price

cat.#

price

cat.#

price

0.25 mm 0.10 µm -60 to 330/430 °C 70205

70208

0.25 µm -60 to 360/430 °C 70220

70223

70226

0.50 µm -60 to 330/360 °C 70235

70238

70241

1.00 µm -60 to 310/340 °C 70250

70253

0.28 mm 0.25 µm -60 to 340/430 °C 70221

70224

70227

0.50 µm -60 to 315/400 °C 70236

70239

1.00 µm -60 to 310/360 °C 70251

70254

70257

3.00 µm -60 to 290/360 °C 70281

70284

0.53 mm 0.25 µm -60 to 340/430 °C 70222

70225

70228

0.50 µm -60 to 330/400 °C 70237

70240

1.00 µm -60 to 310/360 °C 70252

70255

70258

1.50 µm -60 to 300/360 °C 70267

70270

3.00 µm -60 to 290/360 °C 70282

70285

70288

5.00 µm -60 to 270/360 °C 70277

70279

70283

10-Meter

20-Meter

40-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

cat.#

price

0.18 mm 0.20 µm -60 to 325/430 °C 71821

71822

71823

0.40 µm -60 to 315/400 °C 71824

71825

Siloxane (Polysiloxane 20) on MXT®-5

GC_CH00359

Column

MXT®-5, 15 m, 0.28 mm ID, 0.25 µm

(cat.# 70221)

Sample

Polysiloxane 20

Diluent:

Pentane

Injection

Inj. Vol.:

1.0 µL split (split ratio 100:1)

Inj. Temp.:

250 °C

Oven

Oven Temp.:

75 °C to 340 °C at 15 °C/min

Carrier Gas

H

1

, constant flow

Linear Velocity:

40 cm/sec

Detector

FID@ 340 °C

Notes

FID sensitivity: 16 x 10

-11

AFS

MXT® GC Column Ferrule Guide

GC Column ID GC Column OD Ferrule ID

0.18 mm

0.36 ± 0.001

0.4

0.25 mm

0.41 ± 0.001

0.5

0.28 mm

0.56 ± 0.001

0.6

0.32 mm 0.44 ± 0.0015

0.5

0.53 mm

0.74 ± 0.001

0.8