Table of Contents Table of Contents
Previous Page  116 / 628 Next Page
Information
Show Menu
Previous Page 116 / 628 Next Page
Page Background

114

www.restek.com

GC COLUMNS |

METAL MXT® CAPILLARY COLUMNS

Application-Specific Columns: Petroleum & Petrochemical

Simulated Distillation Analysis (C5-C110)

ASTM

Method

D2887

D7213

(D2887-ext)

D5307

D6352

D7096

D7500

D7169

Hydrocarbon

Range

C5–C44

C5–C60

crude

up to C42

C10–C90

gasoline

up to C14

C7–C110

C5–C100

cat. #

70131

70132

70131

70115

70112

70115

70112

70115

70132

10177

70112

70115

70112

70115

Configuration

5 m x 0.53 mm, 0.88 µm

10 m x 0.53 mm, 2.65 µm

5 m x 0.53 mm, 0.88 µm

5 m x 0.53 mm, 0.20 µm

5 m x 0.53 mm, 0.10 µm

5 m x 0.53 mm, 0.20 µm

5 m x 0.53 mm, 0.10 µm

5 m x 0.53 mm, 0.20 µm

10 m x 0.53 mm, 2.65 µm

15 m x 0.53 mm, 5 µm

5 m x 0.53 mm, 0.10 µm

5 m x 0.53 mm, 0.20 µm

5 m x 0.53 mm, 0.10 µm

5 m x 0.53 mm, 0.20 µm

Method Recommended Columns

MXT®-1HT SimDist Column

(Siltek®-treated stainless steel)

(nonpolar phases)

• Stable up to 450 °C—lowest bleed for longest column

lifetime.

• Reliably meets all ASTM D6352, D7169, and D7500

specifications.

• 100% dimethyl polysiloxane phase allows easy

comparisons to historical data.

• Individually tested for guaranteed performance.

• 7

"

coil diameter.

Accurate boiling point determination for medium and heavy fractions using GC

simulated distillation requires columns and phase polymers that are robust enough to

withstand high temperatures without significant degradation.

similar phases

DB-HT SimDis ProSteel, CP-SimDist UltiMetal,

ZB-1X SimDist

5-Meter

10-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.53 mm

0.10 µm

-60 to 430/450 °C

70112

0.20 µm

-60 to 400/430 °C

70115

0.21 µm

-60 to 400/430 °C

70118

0.88 µm

-60 to 380/430 °C

70131

70134

1.00 µm

-60 to 380/400 °C

70130

1.20 µm

-60 to 380/380 °C

70119

2.65 µm

-60 to 360/400 °C

70132

5.00 µm

-60 to 360/400 °C

70133

Hydrocarbons (C5-C106) on MXT®-1HT SimDist at 450 °C

GC_PC1164

Column

MXT®-1HT SimDist, 5 m, 0.53 mm ID, 0.10 µm

(cat.# 70112)

Sample

Custom C5-C106 hydrocarbon standard

Diluent:

Carbon disulfide

Conc.:

1%

Injection

Inj. Vol.:

0.5 µL cold on-column

Temp. Program:

53 °C to 450 °C at 10 °C/min (hold 5 min)

Oven

Oven Temp.:

50 °C to 450 °C at 10 °C/min (hold 5 min)

Carrier Gas

He, constant flow

Flow Rate:

18 mL/min

Detector

FID@ 450 °C

Make-up Gas

Flow Rate:

24 mL/min

Constant Column

+ Constant

Make-up:

42 mL/min

Make-up

Gas Type:

N

2

Data Rate:

20 Hz

Instrument

Shimadzu 2010 GC

Peaks t

R

(min)

1. C5

2. C6

3. C7

4. C8

5. C9

6. C10

7. C11

8. C12

0.938

9. C13

1.586

10. C14

2.425

11. C15

3.365

12. C16

4.332

13. C17

5.290

14. C18

6.217

15. C20

7.966

16. C22

9.566

17. C24

11.051

18. C26

12.426

19. C28

13.689

20. C30

14.897

21. C32

16.035

22. C34

17.110

23. C36

18.133

24. C38

19.108

25. C40 20.096

26. C42

20.923

27. C44

21.759

28. C46

22.556

29. C48

23.317

30. C50 24.051

31. C52

24.752

32. C54

25.422

33. C56

26.079

34. C58

26.701

35. C60

27.305

36. C62

27.878

37. C64

28.439

38. C66

28.975

39. C68

29.499

40. C70 30.002

41. C72

30.489

42. C74

30.906

43. C76

31.414

44. C78

31.862

45. C80 32.294

46. C82

32.719

47. C84

33.132

48. C86

33.529

49. C88

33.927

50. C90 34.310

51. C92

34.689

52. C94

35.059

53. C96

35.423

54. C98

35.773

55. C100 36.120

56. C102 36.463

57. C104 36.793

58. C106 37.118