SRI GCS Literature Detailed ex SRI website 2014 - page 198

Switching / Replacing Traps
Three trapsare includedwithyourSRIPurge&Trap:aTenax
TM
-GR trapandaCarbosieve
TM
trap,both
permanentlypacked,andablank trap. Theblank trapmaybepackedwithanadsorbentof theuser’schoice
or leftblank,dependingon theanalytical situation. Follow the instructionsbelow toaccess the traps for
switchingorreplacement.
1. With theredprotectiveGCcoverraised, remove thePurge&Trapcoverplateby loosening thefourbrass
thumbscrewsat itscorners.
2. Carefullyremove the twosquaresofwhite insulationfromeachvalveovenduct toexpose thefittings that
secure the trapsends to the1/16”O.D. tubing leading to the10portvalve.
3. Gentlyslidethetrapassemblyoutoftheslotsinthevalveovenducts(thereisenoughslackintheheaterand
thermocouplewires topulleither trapabout1 inchoutside theduct).
4. Use twowrenches to loosen the1/8”Swagelok typenuts thatsecure the trapsends tostainlesssteel1/8”-
1/16”reducingunions.
5. The trapheater isaclamshelldesign,consistingof twohalves. Toremove theheaterfrom the trap, loosen
butdonotremovethetwosecuringhexheadscrews. Thetwohalvesoftheclamshellheaterwillopenenough
to let the trapdropout.
6. Attach thereplacement trap to thereducingunionswith the trap’s two1/8”nuts. Usestainlesssteelnuts
andbrass ferruleswhen replacing traps. DONOTusegraphite ferrules, asgraphitehassomeadsorption
propertiesandmay interferewithyouranalysis.
7. Slip the trap into theclamshellheaterand tighten the twohexheadscrews.
8. Gentlypush the trapendsback into theslots in the two interiorductwalls,makingsure that theblack
spacersarebetween theductwallsand the trapheater.
TOAVOIDDAMAGE,ARRANGETHETRAPSSO
THAT ONE TRAP’S HEATERWIRES DONOT LAYACROSS THE OTHER TRAP’S HEATER
.
9. Repeat theprocesswith theother trap ifnecessary. Replace thewhiteduct insulationsquares, thenreplace
andsecure thePurge&Trapcoverplate.
Trapheater
Hex-headscrews
Black
spacer
Black
spacer
Valveovenductwall
(slotted toaccomodate
the trapends)
Heatercartridgewires
Reducing
union
1/8”Swagelok
typenut
1/16”O.D. tubing
leading to thevalve
SRI TrapAssembly
Thermocouplewire (connected to the
hexheadscrewsbyscrew terminals)
INJECTORS