Several criteriawere selected to determine the inertness and
reproducibility of the new injection port design and ensure that proper
vaporization and flow characteristicsweremaintained. The first
experiment was to investigate if injection port discrimination existed in
the new injection port.
To demonstrate the absence of aliphaticmass discrimination theMA
EPHmethod
1
referencemixwas used to determine if the
C
28
toC
20
response ratiomeet the required>0.85.
The test was conducted on anAgilent 5890Series IIGCwith anFID.
The calculated results (Figure 1) show that the response ratio for C
28
to
C
20
is1.0. The response ratiomeets the required standard of >0.85.
1.Method for theDetermination of ExtractablePetroleumHydrocarbons:
Commonwealth ofMassachusetts, May2004, Revision 1.1