12
Semi-Auto Chemistry
MEDICAL ANALYZERS
Features:
Easy Windows operation system with
mouse, large color LCD display.
Support up to 12 kinds of animal:
8 kinds fixed, 4 kinds open.
34 most common tests fixed, 20 more
can be defined by user.
Reagent open system supports flow
cell and cuvette mode, close system
on request.
Model
SACA-11904CV
Principle
Absorbance
Photometric range
-0.500~3.000Abs
Resolution
0.001Abs (displayed), 0.0001Abs (calculated)
Light Source
Halogen lamp
Wavelengths
340, 405, 500, 546, 578, 620, 670nm,
1 more filter optional
Wavelength Accuracy
±1nm
Band Width
8nm
Aspirate Volume
100~300µL
Carryover
<0.1%
Memory
54 test parameters,up to 500 sample results
Interface
RS-232, USB, SD card interface
Flow Cell
25µL Metal-quartz flow cell
Cuvette Mode
12.5mm x 12.5mm cuvette optional
Temperature Control
25°C, 30°C, 37°C, ±0.1°C and ambient temperature
Display
7’’ color LCD(640 x 240, 256 colors)
Input
USB mouse and external keyboard
Output
Build-in printer or external printer (optional)
Net Weight
8KG
Dimensions(mm)
L460×W330×H140
Power supply
AC 110-220V±10%, 50~60Hz
Model
SACA-19100
Principle
Absorbance
Absorbance Range
-0.500 ~ 3.500Abs
Resolution
0.001 Abs (displayed), 0.0001Abs (calculated)
Light Source
Halogen lamp
Wavelengths
Continuously adjusted grating, wavelength range is
330 - 800nm, 1nm by step adjustment
Wavelength Accuracy
± 1nm
Band Width
< 6nm
Flow cell
Metal - quartz flow cell
Temperature Control
25°C,30°C,37°C,±0.1°C and ambient temperature
Carryover
< 1%
Memory
300 test protocols, up to 10000 test results,
support SD to expand memory
Printer
All printers compatible with Windows
Interface
RS-232, USB
Power Supply
AC 100V - 240V ± 10%,50-60Hz
Net Weight
9KG
Dimensions (mm)
L368xW192xH266
SACA-11904CV
SACA-19100
Features:
PC controlled system, Windows XP,
displays the reaction curve in real-time.
Continuously adjustable grating,
enhances the stability and reliability of
result.
SACA-11904CV, Semi-Auto Chemistry, High Quality Analyzer Special For Veterinary Use
SACA-19100, Semi-Auto Chemistry, Grating Inside, PC Controlled
Analytical modes include kinetic, fixed time, bichromatic end point with or without reagent blank or sample blank,
linear or non-linear calibration curves
Real-time curve monitoring enhances the reliability of result
Powerful QC
function: Westguard Multi-rule, Levey-Jennings plot, 2 controls per test
Auto self-check when power on
With lamp
sleeping and wake-up function
Easy software upgrade by SD card, support USB mouse and keyboard
Compact
design enables USER-DIY maintenance
Multi-language software available on request Analytical modes include
kinetic, fixed time, bichromatic end point with or without reagent blank or sample blank, linear or non-linear
calibration curves
Real-time curve monitoring enhances the reliability of result
Powerful QC function: Westguard
Multi-rule, Levey-Jennings plot, 2 controls per test
Auto self-check when power on
With lamp sleeping and
wake-up function
Easy software upgrade by SD card, support USB mouse and keyboard
Compact design enables
USER-DIY maintenance
Multi-language software available on request.
Reagent-open system supports flow cell & cuvette mode, close system on request
Analytical modes include kinetic,
fixed time, bichromatic end point with or without reagent blank or sample blank, linear or non-linear calibration
Powerful QC program: Westguard multi-rule, Levey-Jennings plots, 2 controls per test
Large memory, up to 300 test
protocols can be stored
Multiform result output including patient comprehensive report
Compatible with all
popular external printers connected to PC
Easy communication with PC
Powerful internet connection enables
remote data sharing and OS update
Compact design enables USER-DIY maintenance
Multi-language software
available on request.