Operating Instructions WBL-118 Refrigerated bath and circulator
26
Chapter 5 Features and Specifications
5.1 Key features
SMT assembled and microprocessor-based precise dynamic temperature control with
temperature stability up to ±0.02
°
C;
Independent immersion controller head, DIY for precise temperature system with different
bath capacity and style;
Immersion bath and external circulator in one unit, helps you to save money and make more
space;
Absolute temperature calibration (ATC) for accurate temperature control and AUTO-restart
after power failure for continuous uninterrupted operation;
Designed to be distributed in Europe, North America, Japan, Australia and India.
5.2 Specifications
Models
X
B
WBL-118
B
X
B
X
B
X
B
X
B
Temperature range
°
C -28
+95
Pump flow rate max L/min
12
Pump pressure max mbar
300
Stability
°
C
±0.05~0.1
Setting resolution
°
C
0.1
Display resolution
°
C
0.1
Heater wattage
W
1500
Cooling Capacity@20
BT/RT
W 300
ACC temperature
°
C
40
Bath material
TIG welded, SUS 304
House materials
SPCC steel with powder coating
Liquid connections
Nipple, 1/2”, 12/7mm (od)
Safety class DIN 12876
Class I/NFL, non-flammable heat transfer liquid, No silicone oil
Optional Class II/FL, Class III/FL
EMC for Europe
Class B according to EN 61326-1
EMC for USA and Canada
Class A, for operation only on networks without connected domestic areas
Duty cycle
Continuous
Bath size
mm 170W x350D x 160H
Bath opening
mm 166W x 175D
Overall dimensions
mm 260W x 510D x 607H
Immersion depth
mm
160
Automation and safety
ATC, Temperature sensor failure cutoff and alarm, high-temperature cutoff and
low temperature alarm, blocked pump motor, auto-restart after power failure
and run-dry protection
Power supply
230V~,50/60Hz; 120V~,60Hz; 100V~, 50/60Hz