MRC-2014Hot.svg - page 452

Laboratory Water Purification Systems
WATER-PURIFICATION
Application:
Biochemical analyser inlet, ultra water purification
system inlet. Microbiological media preparation
water, fine chemical, chemical & biochemical
reagent configuration.
Buffer a configuration, photographic flush, ion
chromatography, etc
Feed water:
Potable tap water:
TDS<200ppm, 5-40°C,
1.0 ~ 3.5Kg/cm2.
Model
WPF-90D
Flow rate of DI water
90 liters/hour
Flow rate of RO water
1.5liters/min
Resistivity of DI water
>10 M
/cm
Conductivity of RO wate
< 0.05 x Source Water Conductivity
Granular(>0.2μm)
<1/ml
Microbe
<1cfu/ml
Feed water requirements
tap water, temperature: 5-45°C, pressure: 1.0-4.0Kgf/cm2
Dimension / Weight
L × W × H: 570×600×1500mm / Weight: about 60Kg
Working conditions
Environmental temp.
5°C ~ 35°C
Relative humidity
20% ~ 80%
Electrical Requirements
AC110-220V,50/60Hz; 240W
MRC.7.14
Flow schematic:
LCD Display:
Display
system running status
& various parameters
real-time.
Unique
Consumable
Components:
Integrated desighn of
pretreatment &
subsequent purification
unit.
Built-in Water Tank:
20L airtight plastic
pressure water tank.
Save more
experiment space.
High
Capacity
Pure
Polishing
Resin Cartrige:
22L of pure
polishing resin
cartrige. provides the
highest inorganic ions
removal.
366