Ferrule (b)
Qty. =2
NanoFilter (d)
FilterHolder (c)
Qty. =2
ColumnHolder
Column (e)
CapsuleUnionorNanoFilter (f)
Nut (a)
Qty. =2
C-1210
Column coupler
C-1200
C-1300
CapillarySampleTrapColumns
XX
Packed andunpacked columns
XX
Pressure rated to 5,000psi (345bar)
XX
Direct connect 360µmOD capillary tubing
Upchurch Scientific
®
Capillary Sample TrapColumns are ideal for
separatingandconcentratingand/or purifyingbiological samples.
CapillarySampleTrapColumnAssemblies includeoneormore1µm
NanoFilter
Capsules,eachcontainingeitherastainlesssteel (SST)or
biocompatibletitanium (Ti) frit.Thesweptvolumeoftheseuniquefilters
isonly 10nLeach. Connect your 360µmOD capillary tubingdirectly
toour SampleTrapColumnAssemblies using the fittings provided.
Themaximumflow rate for theseproducts is 10µL/min; 250nL is the
optimal flow rate.
PleaseNote: Reversing flow in these columns can result in loss of
packingmaterial. Formore information regarding theproper setup
involving these Sample TrapColumns, please contact your local
distributor or IDEXHealth&Sciencecustomer service.
Flow
Column
Assembly
Nuts
(a)
Ferrules
(b)
Filter
Holder
Color(c)
NanoFilter
(d)
Column (e)
Capsule
Unionor
NanoFilter (f)
C-1200
P-416 F-152 Tan M-125 (SST) C-1250 (C18)
M-124NF (no frit)
C-1300
P-416BLK F-152 Tan M-126 (Ti)
C-1250 (C18)
M-124NF (no frit)
C-1500
P-416 F-152 Tan M-125 (SST) FS-1000-25
(unpacked)
M-125 (SST)
C-1600
P-416BLK F-152 Tan M-126 (Ti)
FS-1000-25
(unpacked)
M-126 (Ti)
All nuts, ferrules, andcolumnholdersaremadeof PEEKpolymer. Filter holdersaremadeof PEEKpolymer
(non-conductive) or stainless steel (conductive).NanoFilterCapsulebodiesaremadeof PEEKpolymer (non-
conductive) or stainless steel andPEEK (conductive). Seepage162 forNanoFilterCapsulecolor coding.
AbbreviationKey: SST= stainless steel; Ti = titanium; SCX=StrongCationExchange
ComponentsChart
Please refer to thedrawingandpart numbersbelowwhenordering
replacement components:
SPECIFICATIONS&DETAILS
XX
Packingmaterial specifications: C18, high carbon load,
5µm/300A spherical silica andSCX, 5µm 85Amaterial.
XX
Maximum sample loading capacityof 0.1µg and capillarybed
volumeof 0.19µLor less.
RELATEDPRODUCTS
XX
Use theP-116MicroFerrulePlugonpage18 toplug aSampleTrap
Column for storage.
XX
Find360µmODPEEKpolymer and fused silica tubingonpage67.
PartNo.
Description
FritMaterial
Includes
CAPILLARYSAMPLETRAPCOLUMNS
Assemblies
C-1200
C18Column
SST
(1) 2-pkC-1250, (1)M-125,
(2) P-416, (2) F-152, (1)M-124NF
C-1300
C18Column
Ti
(1) 2-pkC-1250, (1)M-126,
(2) P-416BLK, (2) F-152, (1)M-124NF
C-1500
UnpackedColumn
SST
(1) FS-1000-25, (2)M-125,
(2) P-416, (2) F-152
C-1600
UnpackedColumn
Ti
(1) FS-1000-25, (2)M-126,
(2) P-416BLK, (2) F-152
ColumnCoupler andReplacement Parts
PartNo.
Description
Swept
Volume
Qty.
C-1210
ColumnCoupler, PEEK
ea.
C-1250
C18Columns, 100µm IDx2.5cmx360µmOD
2-pk
F-152
MicroFerrule for 360µmOD tubing, PEEK
ea.
FS-1000-25
UnpackedColumn, 100µm IDx2.5cmx360µmOD —
ea.
M-124NF
CapsuleUnion, noFrit, PEEK
9.5nL
ea.
M-125
1µmNanoFilterCapsules,withSSTFrits
10nL
2-pk
M-126
1µmNanoFilterCapsules,withTi Frits
10nL
2-pk
P-416
FemaleNut,Natural PEEK
ea.
P-416BLK
FemaleNut,BlackPEEK
ea.
P-416G
FemaleNut,GreenPEEK
ea.
FILTERS&COLUMNACCESSORIES