SRITN12 - page 287

POPULAR CONFIGURATION GCs
Multiple Gas Analyzer #2
System Overview
TheSRIMultipleGasAnalyzer#2(MG#2) isconfiguredon the8610Cchassis. It isequippedwithagas
samplingvalveplumbedwithdualsampleloopsinaheatedvalveoven,andtwopackedcolumnsinthecolumn
oven. Thebasicmodel, shownbelow,comeswithaTCDdetector. TheMG#2mayoptionallybeequipped
withaFID/methanizerorHIDdetectors inaddition to theTCD. Acapillarycolumn inparallelwith the
Hayesep-Dcolumn isanoptionforseparatingouthydrocarbons throughC
20
.
TheMG#2separatesawidevarietyofpeakswithoutco-elution
by turning thecarriergas flow to the twopackedcolumnsONand
OFFindividuallyatdifferenttimesduringtherun. Thecarriertothe
Molecular Sieve 13x column (carrier #1) is turnedON first to
complete theseparationofH
2
,O
2
,N
2
,CH
4
andCO. At thispoint,
theMoleSievecarrierflowisturnedOFFandtheHayesep-Dcarrier
(carrier#2) is turnedON. All compounds in theC
1
-C
6
rangeare
thenseparatedby theHayesep-Dcolumn. TheMoleSievecolumn
isconnected tosample loop1, and theHayesep-D to loop2.
MoleSieve
(loop 1)
Hayesep-D
(loop 2)
Hayesep-Dcolumn
Gas sampling valve
with twosample loops
TCDdetector
On-column injector
Samplestream
InandOut
Valveoven
Columnoven
MolecularSieve13xcolumn