SRITN12 - page 295

POPULAR CONFIGURATION GCs
Mud-Logger
General Operating Procedure Part 1:
Total Hydrocarbons Using the CCD Detector
1. Connectzerogas tosample inletat10psi. Zerogashasnohydrocarbons.
2. Zero theCCDdetectorsignalusing theAutoZerobuttonfor itschannel (typicallychannel2).
3. Connectcalibrationgasstandard to thesample inletat10psi. Calibrationgas is typically100%methane.
4. TheCCDsignalwillincreaseapproximately300millivoltswhilerunning100%methane.
5. InPeakSimple,open theCCDChannelDetailsdialogboxbyright-clicking in thatchannel’schromato-
gramwindow. Enter thegainfactorwhichwillmultiply the300millivoltsignal toproduce thedesired
concentrationunit. Forexample:
300x .33=100 if thedesiredunit ispercent.
300x3333=1,000,000 if thedesiredunit ispartspermillion
6. Also in theChannelDetailsdialogbox,selectDataLoggermodebyclicking in theappropriate
checkbox. TheCCDsignal times thegainfactorwillbedisplayedon thescreen in largenumbers.
PeakSimple
ChannelDetailsdialogbox
Chromatogramwith
channel 2 inDataLoggermode
1...,285,286,287,288,289,290,291,292,293,294 296,297,298,299,300,301,302,303,304,305,...505
Powered by FlippingBook