9

blesome compounds at 0.5 ng on column relative

to the method minimum, the average RF for the

calibration range (0.5-120 ng on-column), and lin-

earity evaluated by RRF RSD and linear regression.

Calibrations were also assessed for the full list

of compounds. For the initial calibration (ICAL)

as a whole to meet acceptance criteria, less

than 10% of the individual compounds may

have failing RSDs (or correlations, if alterna-

tive fit methods are used). When the peak

response RSDs were evaluated over the entire

calibration range for the full list of compounds,

the average RSD was 8.7% and only 10 of the

compounds tested had RSDs greater than

20%. Linearity results for both indicator and

non-indicator compounds demonstrate that

detection limits can be lowered for semivola-

tiles analysis by using a highly inert system that

allows the lower end of the calibration range to

be extended.

Figure 1:

Extend the calibration range for difficult semivolatiles down to 0.5 ng on-column by using a highly inert analytical system.

(total ion chromatogram of EPA Method 8270 and Appendix IX compounds)

www.restek.com

|

800-356-1688

| Feedback? E-mail

advantage@restek.com

Restek Recommends

For ultimate sample path inertness:

EZ Twist Top® Injection Port

www.restek.com/eztwist

Sky™ Inlet Liners

www.restek.com/sky

Flip Seal™ Inlet Seals

www.restek.com/flip

Rxi®-5Sil MS Columns

www.restek.com/rxi

Figure 2:

Response differential for

2,4-dinitrophenol and 4-nitrophenol.

1. 2,4-Dinitrophenol (m/z 184)

2. 4-Nitrophenol (m/z 139)

See Figure 1 for analytical conditions.

45,46

4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 6.20 6.40 6.60 6.80 7.00 7.20 7.40 7.60 7.80 8.00 8.20 8.40 8.60 8.80 9.00 9.20 9.40

12

13,14

15

16,17

18

19

20,21

22 2324

25

26,27,28

29,30,31

32

33

34,35

36

37

38

39

40

41

42,43

44 47,48

49

50

51

52,53 54

55

56

57

58,59

60

61 62

63

64

65

66

67

69

70

71

72

73

74

75

76

77

78,79

80

81

82 83,84

85,86

87

8889

90,91

92,93

94

95

96

97

98

99,100

101

102103

104

105

106

107

108,109

110

111

68

4.40

1,2

3

4 5

6

7 8

9

10 11

112

113

114

115

116

117

118,119 121

120

122

123

124

125,126

127,128

129

131

131,132,133

134

135

136,137

138

139

140

141

142

143

144 145

2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50

7.50 8.00 8.50 9.00 9.50 10.00 10.50 11.00 11.50 12.00 12.50 13.00 13.50 14.00 14.50 15.00 15.50 16.00 16.50 17.00

7.00

EIC:

12 = m/z 106

15,16 = m/z 93

EIC: m/z 162

EIC: m/z 232

EIC:

m/z 106

EIC:

143 = m/z 276

144 = m/z 278

Time (min)

4.60

4.40

4.80

12

15

16

14.00

137 138

16.50

143 144

7.20

66

67

7.25

8.05 8.10

87

86

GC_EV1269

Column

: Rxi®-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623);

Sample

: 8270 MegaMix® (cat.# 31850), 8270 Benzidines mix (cat.# 31852), Benzoic acid (cat.# 31879), Revised B/N surrogate mix (cat.# 31888), Acid sur-

rogate mix (4/89 SOW) (cat.# 31063), Revised SV internal standard mix (cat.# 31886), Appendix IX mix #1 (cat.# 31625), Appendix IX mix #2 (cat.# 31806); Diluent: Dichloromethane; Conc.: 0.5 µg/mL (IS/SS 20 μg/mL);

Injection:

1 µL pulsed splitless (hold 0.59 min); Liner: Sky™4 mm single taper w/wool (cat.# 23303); Inj. Temp.: 270 °C; Pulse Pressure: 30 psi (206.8kPa); Pulse Time: 0.64 min; Purge Flow: 100 mL/min;

Oven:

40 °C (hold 1

min) to 280 °C at 25 °C/min to 320 °C at 5 °C/min (hold 1 min);

CarrierGas

: He, constant flow; Flow Rate: 1.2 mL/min;

Detector

: MS; Mode: Scan; Transfer Line Temp.: 280 °C; Analyzer Type: Quadrupole; Source Temp.: 276 °C;

Quad Temp.: 150 °C; Electron Energy: 70 eV; Solvent Delay Time: 2.19 min; Tune Type: DFTPP; Ionization Mode: EI; Scan Range: 35-550 amu; Scan Rate: 5.36 scans/sec;

Instrument:

Agilent 7890A GC & 5975C MSD;

Notes:

7890 Siltek®-treated EZ Twist Top® split/splitless injection port (cat.# 22178), Flip Seal™ dual Vespel® ring inlet seal (cat.# 23411); For peak identifications, visit

www.restek.com

and enter GC_EV1269 in the search.

• Quantify down to 0.5 ng on-column

• Separate key PAHs.

• Analyze 145 semivolatiles in 17 minute analysis.

For more environmental applications, visit

www.restek.com/enviro

3

4

5

6

7

8

9

10

7.30 7.40 7.50 7.60 7.70 7.80 7.90 8.00 8.10 8.20 8.30 8.40 8.50

1

2

Time

1

2

1

2

0.5 ng

On-Column

10 ng

On-Column

2,4-DNP

RF=0.055

2,4-DNP

RF=0.19