Figure 2

Stx

-CLPesticides and Stx

-CLPesticides2 columns provide rapid, excellent separation of chlorinated pesticides, and a Siltek

-deactivated Drilled

Uniliner

®

inlet liner helps ensure high responses for sensitive analytes.

• 12 •

www.restekcorp.com

by Gary Stidsen, Innovations Manager

Inert Inlet System Improves

Responses for Chlorinated

Pesticides

Using the Drilled Uniliner

®

Inlet Liner

Inert sample path eliminates injection port discrimination.

Reduce detection limits by using a splitless injection port without an on-column

injector.

For years, chemists analyzing chlorinated pesticides

have tried many different injection techniques in

attempts to find the best balance between inertness

and ability to contend with sample contamination.

Cool on-column, split, splitless, and direct injec-

tion, and variations of these injection techniques,

are used today.

Now, a specially modified injection port liner, devel-

oped by Restek chemists, reduces sample contact

with active metal parts in split/splitless injection ports.

This Drilled Uniliner

®

liner, shown in Figure 1, gives

the benefits of both direct injection and splitless

injection. The advantage of this liner is that the col-

umn is connected to the liner by a press-fit connec-

tion, thus preventing the sample from contacting the

metal at the bottom of the injection port. Also, the

hole on the side of the liner allows the purge flow to

escape from the liner when the injection mode is

switched from splitless to split.

Inertness

With the Drilled Uniliner

®

liner, the sample is trans-

ferred directly from the injection port to the col-

umn and contacts only this glass inlet liner. The

configuration allows the sample to be "funneled"

into the column entrance, thereby eliminating the

need for vaporization aids such as fused silica

wool. The test probes endrin and 4,4'-DDT are

good indicators of injection port inertness. Analyses

of the performance evaluation mixture (PEM) show

low breakdown of these compounds when using

this liner.

For maximum inertness, we recommend using the

Drilled Uniliner

®

liner in combination with an Stx

-

CLPesticides capillary column, as shown in Figure

2. Siltek

surface deactivation in these columns fur-

ther ensures maximum responses for the labile pes-

ticides endrin, 4,4'-DDT, and methoxychlor. A spe-

cial polymer formulation designed for organochlo-

rine pesticides enables you to achieve excellent sep-

aration of the 22 chlorinated pesticides in Figure 2

in less than 15 minutes. Comparable analyses on

Rtx

®

-CLPesticides columns take up to 24 minutes

(Figure 2, inset).

Calibration

Linearity and continuing calibration checks for the

chlorinated pesticides also are critical parameters

that must be monitored. Table 1 indicates typical

1. 2,4,5,6 tetrachloro-

m

-xylene (IS)

2.

α

-BHC

3.

γ

-BHC

4.

β

-BHC

5.

δ

-BHC

6. heptachlor

7. aldrin

8. heptachlor epoxide

9.

γ

-chlordane

10.

α

-chlordane

11. 4,4'-DDE

12. endosulfan I

13. dieldrin

14. endrin

15. 4,4'-DDD

16. endosulfan II

17. 4,4'-DDT

18. endrin aldehyde

19. methoxychlor

20. endosulfan sulfate

21. endrin ketone

22. decachlorobiphenyl (IS)

Oven temp.:

110°C (hold 1 min.) to 245°C @ 20°C/min.

to 300°C @ 6°C/min.

Inj. & det. temp.:

210°C / 310°C

Carrier gas:

helium

Dead time:

0.8min. @ 120°C

Inlet liner:

Siltek

Drilled Uniliner

®

liner (cat.# 21055-214.5)

Inj.:

1µL direct injection of 20/40/200ng/mL std.

concentration in hexane

Make-up gas:

nitrogen

1

2 3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

21

22

30m, 0.32mm ID, 0.5µm

Stx

-CLPesticides

(cat.# 11544)

30m, 0.32mm ID, 0.25µm

Stx

-CLPesticides2

(cat.# 11444)

1

2

3

4

5

6

7

8

9

10

12

11

13

14

15

16

1718

19

20

21

22

GC_EV00512

Figure 1

The drilled hole in a Uniliner

®

injection port

liner makes direct injection possible with

EPC systems by equalizing pressure in the

injection port.

8

10

12

14 min.

8

10

12

14 min.

6

9

10

11

12

13

14

15

16

17

18

19

20

1

2

3

4

5

6

7

8

21

22

min. 10

24

GC_EV00075

On-column concentration:

16–160pg organochloride

pesticide mix AB#2

(cat.# 32292)

Oven temp.:

120°C (hold 1 min.) to 300°C

(hold 10 min.) @ 9°C/min.

Inj. port:

Direct, Uniliner

®

liner

(cat.# 20335), at 200°C

Detector:

ECD, 300°C with Anode Purge

Dead time:

1.9 min.

Head pressure:

8.7psi (constant)

Flow rate:

1.3mL/min. @ 120°C, helium.

30m, 0.32mm ID, 0.25µm

Rtx

®

-CLPesticides2

(cat.# 11324)

Traditional analyses on

Rtx

®

-CLPesticides columns take

up to 24 minutes.