69

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Mineral Oil and Motor Oil on Rtx®-Mineral Oil

Mineral Oils/Extractable Petroleum Hydrocarbon Analysis

GC_PC1169

Column

Rtx®-Mineral Oil, 15 m, 0.32 mm ID, 0.15 µm (cat.# 18074)

using IP deactivated guard column 2 m, 0.53 mm ID (cat.# 10047)

Sample

Custom mineral oil/motor oil mix

Diluent:

Hexane

Conc.:

500 µg/mL

Injection

Inj. Vol.:

0.5 µL cold on-column

Temp. Program:

53 °C to 300 °C at 10 °C/min (hold 20 min)

Oven

Oven Temp.:

50 °C to 300 °C at 10 °C/min (hold 20 min)

Carrier Gas

H

2

, constant flow

Linear Velocity:

40 cm/sec @ 50 °C

Dead Time:

0.625 min @ 50 °C

Detector

FID@ 330 °C

Make-up Gas

Flow Rate:

30 mL/min

Make-up

Gas Type:

N

2

Data Rate:

20 Hz

Instrument

Agilent/HP6890 GC

Notes

Black trace = mineral oil

Green trace = motor oil

Red trace = C10-C40 standard

Fused Silica Capillary & PLOT Column

Ferrule Guide

Rtx®-Mineral Oil Columns

(fused silica)

• Application specific columns meet DIN EN ISO 9377-2:2000 requirements.

• Optimized column dimensions for fast mineral oil screening.

• Surface linked phase guarantees long lifetime, robustness, and stability to 400 °C.

The Rtx®-Mineral Oil stationary phase and column dimensions were optimized for the

fast screening of mineral oils in extracts from solids and water samples according to

DIN EN ISO 9377-2:2000. The 0.10 µm column is the gold standard for the method,

whereas the 0.15 µm column provides more complete separation of C10 from the sol-

vent peak when large injection volumes are used. Compared with common industry

solutions, the unique surface bonding of the Rtx®-Mineral Oil column ensures long

column lifetime, even at higher temperatures. These unique columns can be used at

temperatures ranging from 380 °C (isothermal) to 400 °C (programmable), and each

column is tested individually for bleed to ensure exceptional performance at these ex-

treme conditions.

similar phases

Select Mineral Oil

15-Meter

ID

df

temp. limits

cat.#

price

0.32 mm 0.10 µm -60 to 380/400 °C

18079

0.15 µm -60 to 380/400 °C

18074

0.30 µm -60 to 380/400 °C

18075

GC Column ID

Ferrule ID

0.15 mm

0.4

0.18 mm

0.4

0.25 mm

0.4

0.32 mm

0.5

0.53 mm

0.8

Restek’s state-of-the-art facility and rigorous

product testing programs ensure you get the