Page Background www.restek.com

67

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Dioxin & Furan Analysis

Dioxins (TCDDs) and Furans (TCDFs) in Fly Ash on an Rxi®-5Sil MS column

GC_EV1194

TCDDs

GC_EV1195

Column

Rxi®-5Sil MS , 60 m, 0.18 mm ID, 0.10 µm (cat.#43607)

Sample

Fly ash extract

Diluent:

Nonane

Injection

Inj. Vol.:

1 µL splitless

Liner:

2 mm splitless liner (cat.# 20712)

Oven

Oven Temp.:

120 °C (hold 1 min) to 160 °C at 10 °C/min to 300°C at 2.5°C/min

Carrier Gas

He, constant flow

Flow Rate:

1 mL/min

Detector

Waters AutoSpec Ultima Mass Spectrometer

Source Temp.:

280 °C

Ionization Mode: EI

Notes

Electron Ionization at 40eV

Acknowledgement

Chromatogram courtesy of Karen MacPherson,

Li Shen, Terry Kolic, and Eric Reiner at the

Ontario Ministry of the Environment

TCDFs

Similar to: (5%-phenyl)-methylpolysiloxane

Rxi®-5Sil MS Columns

(fused silica)

(low-polarity phase; Crossbond® 1,4-bis(dimethylsiloxy)phenylene dimethyl polysiloxane)

• Engineered to be a low-bleed GC-MS column.

• Excellent inertness for active compounds.

• Ideal for use in dual column confirmation of dioxin and furan.

• Temperature range: -60 °C to 350 °C.

The Rxi®-5Sil MS stationary phase incorporates phenyl groups in the polymer back-

bone. This improves thermal stability, reduces bleed, and makes the phase less prone

to oxidation. Rxi®-5Sil MS columns are ideal for GC-MS applications requiring high

sensitivity, including use in ion trap systems.

Rxi®-5Sil MS Structure

95%

5%

CH

3

CH

3

Si

O

O

CH

3

CH

3

Si

CH

3

CH

3

Si

similar phases

DB-5ms, DB-5msUI, VF-5ms, ZB-5ms,

ZB-SemiVolatiles, Rtx-5Sil MS

30-Meter

60-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.18 mm 0.10 µm -60 to 320/350 °C

43607

0.25 mm 0.25 µm -60 to 320/350 °C

13623