GC DETECTORS
Reduction Gas Detector - RGD
Maintenance
ReactionTubePacking andReplacement
WARNING!
Alwaysavoidcontactwith theHgO. Wearprotectiveglovesandeyecovers,and takecarenot
to inhale thepowder.
Remove thequartzwoolandHgOwitha longsharpobject,anddisposeof themproperly. WipeanyHgOoff
thesharpobjectwithadisposablewipeor tissueanddiscard. Re-pack thereactor tubewith1/8”ofHgOand
addfreshquartzwool. TheHgOmustcover thefaceof thefritentirely toforce thesample through theHgO
and into theUVcell.
Thegraphiteferruleseals theendof thereaction tubearound thecolumn. Change thegraphiteferrulewhen
youchange thereaction tube.
Graphite ferrule Column
Quartzwool
Frit
1/8”HgO
2”
SRIprovidesyouwithonereplacementreactor tube inyouraccessorieskit. However,youmayalsoremove
themercuricoxide(HgO)andre-pack thereactor tube, forwhichyouwillneedfreshHgOandquartzwool.
HgOpowder isavailable fromSigma-Aldrich:part#203793-2G:www.sigmaaldrich.com,orcall:1-800-
325-5832. Quartzwool isavailablefromRestek:catalog#20999:www.restekcorp.com,orcall1-800-356-
1688.
Toremove thereactor tube, loosen thefitting thatsecures it to thecolumnandRGD jet,andpull thereactor
tubeout.
Loosen this fitting