RtxPresentations07 - page 396-397

Analysis
Sample chromatograms are shown for the separation of several PCBs and dioxins by chlorination level. PCB data for the
analysis of two referencematerials (sediment and biota) are summarized in the tables below. Thismethodwas also evaluated
for three different extracts, soil, biota, and air versus the conventional separation on the 60M column. This data appears as a
comparison of the “fast”method to the “conventional”method in the table.
Analysis of PCBCongenerReference Sediment EC-3
PCBCongener
ExpectedValue ng/g
N
AverageValue
18
9.0+/- 4.7
8
9.2
28
18.6+/- 8.6
8
14.7
52
35.6 +/- 12.9
8
28.9
105
13.1 +/- 4.3
8
18.9
118
28.5 +/- 5.4
8
28.6
138
25.2 +/- 6.3
8
26.0
153
24.2 +/- 4.1
8
22.6
170
8.9+/- 1.3
8
9.6
180
15.4 +/- 6.6
8
13.7
Extract
Compound
Conventional GC/HRMS (60M)
Fast GC/HRMS (40M)
Soil
2,3,7,8-T
4
CDD
2.2 ppt
2.3 ppt
Fish
2,3,7,8-T
4
CDD
14 ppt
13 ppt
Air
1,2,3,7,8-P
5
CDD
0.093 pg/M
3
0.093 pg/M
3
1...,376-377,378-379,380-381,382-383,384-385,386-387,388-389,390-391,392-393,394-395 398-399,400-401,402-403,404-405,406-407,408-409,410-411,412-413,414-415,416-417,...
Powered by FlippingBook