2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00
0
200000
400000
600000
800000
1000000
1200000
1400000
1600000
1800000
2000000
2200000
2400000
2600000
Time-->
Abundance
TIC:A3021764.D
Rtx-624
The 30mx 0.25mm ID columnwas unable to
adequately resolveTBA fromMTBE.
Extracted ion chromatograms illustrate that a
minor ionofMTBE (59) contributed to the area
ofTBA.. The optimizedoven conditionswere as
follows: 35°C (hold 6min.) to 90°C@ 8°C/min.
(hold 1min.) to 220°C@ 16°C/min. (hold 2
min.). TheGC runtimewas 24minuteswitha
cycle time of 30minutes. SinceTBA is a
breakdown component ofMTBE these column
dimensions and conditions arenot
recommended for the analysis ofTBA.