3
Trifluralin
Figure2.
Ion Ratio for a group of pesticide standards in a Calibration Table. The Trifluralinmass
spectrum demonstrates how two substantial ions (264 and 306) have been chosen as the Ion
Masses with an Expected Ion Ratio of 1.22.
Figure3.
Peak Table for a spinach extract showing results of a check of Ion Ratios for the spiked
pesticides. Note that the Quant Masses and the IonMasses overlap inmost cases. If Ion Ratio
Result were "Failed" instead of "Passed", this would indicate a possible bias in quantification. Note
that the Expected Ion Ratio could also be displayed in the Peak Table.