Resolution new MXT-1HT Sim Dist

5m x 0.53 x 0.2um

C50/C52

R = 2.45

Column : MXT®-1HT Sim Dist, 5m, 0.53mm ID, 0.20µm

Instrument: : Shimadzu2010

Sample :C5-C100,1% in carbondisulfide;

Injection : 1µL viaPTV; Inj. temp.:53°C to430°C@ 10°C/min. (hold5min.);

Carrier gas :He, constant flow; Flow

rate:18mL/min.

;

Oven temp. : 50°C to430°C @ 10°C/min. (hold5min.);

Det.

:FID @ 430°C;

0

10

20

30

40

Time (min)

0

10

20

30

40

Time (min)

Varian

Restek

J&W / Agilent

Comparison of Resolution values

5m x 0.53 x 0.2um

5m x 0.53 x 0.17um

5m x 0.53 x 0.15um

C50/C52

R = 1.18

C50/C52

R = 2.45

C50/C52

R = 2.25

NOT OK !

OK

OK

0

10

20

30

40

Time (min)

Resolution values

The higher resolution value indicates the Restek

MXT-1HT Sim Dist:

• Has a smoother coated surface of the PDMS

phase over the length of the column

• The higher resolution value allows for a longer

column lifetime