712
WATER QUALITY
W
Turbidity
TU-2016, Turbidity Meter
Designed to meet ISO 7027.
NTU (Nephelometric TURBIDITY Unit) measuring unit.
Wide and auto measurement range : 0 to 1,000 NTU.
High resolution : 0.01 NTU/1 NTU.
The unique optics structure, enables the instrument to read low value of
TURBIDITY to the high level up to 1,000 NTU.
Four operation buttons and two calibration points, easy operation.
Jumbo LCD, easy readout.
Microprocessor circuit assures maximum possible accuracy, provides
special functions and features.
Battery operated for field and on-site testing convenience.
Data hold function for freezing the desired value on display.
Records Maximum and Minimum readings with Recall.
Heavy duty & compact housing with hard carrying case, designed for easy
carry out & operation.
Auto shut off is available to save battery life.
Application : Test municipal water, food and beverage water, or other
aqueous solution where fluid clarity is important.
Model
TU-2016
Circuit
Custom one-chip of microprocessor LSI circuit.
Display
LCD size : 41 mm x 34 mm
Range
0.00 to 50.00 NTU, 50 to 1,000 NTU * NTU : Nephelometric Turbidity Unit * Auto range
Resolution
0.01 NTU/ 1 NTU
Accuracy
± 5 % F.S. or ± 0.5 NTU, which ever is greater.
Light source
LED, 850 nm.
Detector
Photo diode
Standard
Meet ISO 7027.
Response time
Less than 10 seconds.
Sample volume needed
10 mL.
Data Hold
Freeze the display reading.
Memory Recall
Maximum & Minimum value.
Sampling Time
Approx. 1 second.
Power off
Auto shut off saves battery life or manual off by push button.
Calibration points
0 NTU, 100 NTU.
Operating Temperature
0 to 50°C.
Operating Humidity
Less than 85% R.H.
Power Supply
DC 1.5 V battery ( UM4, AAA ) x 6 PCs, or equivalent.
Power Current
Stand by
Approx. DC 3.5 mA.
Testing
Approx. DC 36 mA.
Weight
320 g/0.70 LB. @ Battery is included.
Dimension
155 x 76 x 62 mm ( 6.1 x 3.0 x 2.4 inch)
Accessories Included
* Testing bottle with 0 NTU standard solution: 1 PC
* Testing bottle with 100 NTU standard solution: 1 PC * Empty testing bottle: 2 PCs
* Clean cloth: 1 PC * Clean solution (Distill water): 1 bottle* Hard carrying case, CA-08: 1 PC
Optional Accessories
* Testing bottle with 100 NTU standard solution, TU-100NTU
* Testing bottle with 0 NTU standard solution, TU-0NTU * Empty testing bottle, 0601
TU-2016