MRC-2013-Catalog - page 703

703
WATER QUALITY
W
PH/ORP Data Logger
Model
PH-230SD
Circuit
Custom one-chip of microprocessor LSI circuit.
Display
LCD size : 52 mm x 38 mm LCD with green backlight ( ON/OFF ).
Measurement Function
pH, ORP (mV)
Datalogger
Sampling Time
Setting range
Auto
1 sec to 8 hour 59 min. 59 sec.
@ For anemometer measurement, the sampling time setting value should
be
2 seconds.
@ Sampling time can set to 1 second, but memory data may loss.
Manual
Push the datalogger button once will save data one time.
@ Set the sampling time to 0 second.
@ Manual mode, can also select the 1 - 99 position (Location) no.
Memory Card
SD memory card. 1 GB to 16 GB.
Advanced setting
* SD memory card Format * Set clock time ( Year/Month/Date, Hour/Minute/ Second ) * Set
sampling time * Auto power OFF management * Set beep Sound ON/OFF * Decimal point of
SD card setting * Set temperature unit to °C or °F * Set pH manual Temp. compensation value
Data Hold
Freeze the display reading
Memory Recall
Maximum & Minimum value
Sampling Time of Display
Approx. 1 second
Data Output
RS 232/USB PC computer interface. * Connect the optional RS232 cable UPCB-02 will get the
RS232 plug * Connect the optional USB cable USB-01 will get the USB plug.
Power Supply
* Alkaline or heavy duty DC 1.5 V battery (UM3, AA) x 6 PCs, or equivalent
* DC 9V adapter input. (AC/DC power adapter is optional)
Power Current
Normal operation ( w/o SD card save data and LCD Backlight is OFF) : Approx. DC 15 mA.
When SD card save the data and LCD Backlight is OFF) : Approx. DC 37 mA.
Operating Temperature
0 to 50°C
Operating Humidity
Less than 85% R.H.
Weight
489 g/1.08 LB.
Dimension
177 x 68 x 45 mm (7.0 x 2.7x 1.9 inch)
Accessories Included
* Instruction manual: 1PC
Optional Accessories
* PH electrode PE-03, PE-11, PE-01, PE-02, PE-07, PE-06HD, PE-04HD, PE-05T, PE-03K7 * ATC
Temp. probe TP-07 * pH 7 buffer solution PH-07 * pH 4 buffer solution PH-04 * ORP Electrode
ORP-14 * Hard carrying case CA-06 * Soft carrying case CA-05A * SD memory card ( 1 GB ) *
SD memory card ( 2 GB ) * AC to DC 9V adapter. * USB cable, USB-01. * RS232 cable, UPCB-02.
* Data Acquisition software,SW-U801-WIN.
pH : 0 to 14.00 pH, mV (ORP) : ± 1999 mV.
pH measurement can select
ATC or manual Temp. adj.
Optional ATC probe for pH measurement.
pH
measurement can make the auto calibration for pH 7, pH 4 and pH 10 or
other value.
Real time SD memory card Datalogger, it Built-in Clock and
Calendar, sampling time set from 1 sec to 8 hour 59 min. 59 sec.
Manual
datalogger is available (set the sampling time to 0), during execute the
manual datalogger function, it can set the different position (location) No.
(position 1 to position 99).
Innovation and easy operation, computer is
not need to setup extra software, after execute datalogger, just take away
the SD card from the meter and plug in the SD card into the computer, it
can down load the all the measured value with the time information (year/
month/date/hour/minute/second) to the Excel directly, then user can make
the further data or graphic analysis by themselves.
SD card capacity : 1
GB to 16 GB.
LCD with green light backlight, easy reading.
Can default
auto power off or manual power off.
Data hold, record max. and min.
reading.
Microcomputer circuit, high accuracy.
Power by UM3/AA (1.5V)
x 6 batteries or DC 9V adapter.
RS232/USB PC COMPUTER interface.
Wide
applications: water conditioning, aquariums, beverage, fish hatcheries, food
processing, photography, laboratory, paper industry, plating industry, quality
control, school & college, water conditioning.
PH-230SD, PH meter, SD Card, Real time Datalogger, Patent + ORP
Electrical Specifications (23±5
°
C):
PH Electrode
Optional, Any PH electrode with BNC connector.
Measurement
PH
0 to 14 PH
mV
-1999 mV to 1999 mV
Input Impedance
10^12 ohm
Temperature
Compensation for pH
measurement
Manual
0 to 100°C, be adjusted by push button on front panel.
Automatic (ATC)
With the optional temperature probe ( TP-07 ) 0 to 65°C.
pH Calibration
PH7, PH4, and PH10, 3 points calibration ensure the best linearity and accuracy.
Measurement
Range
Resolution
Accuracy
PH
0 to 14 PH
0.01 PH
±(0.02 PH+2d)
mV (ORP)
0 to 1999 mV
1 mV
±(0.5%+2d)
* PH accuracy is based on calibrated meter only.
PH-230SD