MRC-2013-Catalog - page 700

700
WATER QUALITY
W
PH/ORP/DO/CD/TDS/Salt, Data Logger
Conductivity probe
Optional, Carbon rod electrode for long life.
Function
* Conductivity (uS, mS) * TDS (Total Dissolved Solids, PPM) * Temperature (°C, °F)
Temperature Compensation
Automatic from 0 to 60
°C
(32 - 140
°F
),
with temperature
compensation factor variable between 0 to 5.0% per C.
Probe Operating Temp.
0 to 60 °C
Probe Dimension
Round, 22 mm Dia. x 120 mm length.
Optional probe & accessories
* Conductivity probe CDPB-03 * 1.413 mS Conductivity Standard Solution CD-14
Conductivity probe
Optional, Carbon rod electrode for long life.
Measurement Range/Resolution
0 to 12 % salt (% weight) / 0.01 % salt.
Accuracy
0.5 % salt value * F.S.: full scale.
Temperature Compensation
Automatic from 0 to 60°C (32 - 140°F), with temperature compensation
factor variable between 0 to 5.0% per C.
Probe Dimension
Round, 22 mm Dia. x 120 mm length.
Optional probe & accessories
* Salt probe (Conductivity probe) CDPB-03
Conductivity
Function
Measuring Range
Resolution
Accuracy
°C
0°C to 60°C
0.1 °C
± 0.8 °C
°F
0°F to 140°F
0.1 °F
± 1.5 °F
@ 23± 5°C
Range
Measurement
Resolution
Accuracy
200 uS
0 to 200.0 uS
0.1 uS
±(2% F.S.+1d)
* F.S. - full scale
2 mS
0.2 to 2.000 mS
0.001 mS
20 mS
2 to 20.00 mS
0.01 mS
200 mS
20 to 200.0 mS
0.1 mS
* Temperature Compensation: Automatic from 0 to 60
°C
(32 - 140
°F
), with temperature compensation factor variable
between 0 to 5.0% per C. * The accuracy is specified under measurement value
100 mS. * mS - milli Simens * @ 23± 5
°C
Range
Measurement
Resolution
Accuracy
200 PPM
0 to 132 PPM
0.1 PPM
±(2% F.S.+1d)
* F.S. - full scale
2,000 PPM
132 to 1,320 PPM
1 PPM
20,000 PPM
1,320 to 13,200 PPM
10 PPM
200,000 PPM
13,200 to 132,000 PPM
100 PPM
* Temperature Compensation: Automatic from 0 to 60
°C
(32 - 140
°F
), with temperature compensation factor variable
between 0 to 5.0% per C. * The accuracy is specified under measurement value
100 mS. * mS - milli Simens * @ 23±5
°C
Temperature
TDS (Total Dissolved Solids)
Conductivity (uS, mS)
Salt
Dissolved oxygen
Oxygen Probe
Optional, The polarographic type oxygen probe with
Measurement & Range
Dissolved Oxygen
0 to 20.0 mg/L (liter).
Oxygen in Air
0 to 100.0 %.
Temperature
0 to 50 °C.
Resolution
Dissolved Oxygen
0.1 mg/L.
Oxygen in Air
0.1 % O2.
Temperature
0.1 °C.
Accuracy (23± 5
°
C)
Dissolved Oxygen
± 0.4 mg/L.
Oxygen in Air
± 0.7% O2.
Temperature
± 0.8 °C/1.5 °F.
Probe Compensation & Adj.
Temperature
0 to 50 °C, Automatic
Salt
0 to 50 % Salt
Height ( M. T.)
0 to 8900 meter
Probe Weight
335 g/0.74 LB ( batteries & probe included )
Probe Size
190 mm x 28 mm Dia. ( 7.5” x 1.1” Dia. )
Optional Accessories
* Oxygen probe OXPB-11 * Spare Probe head with Diaphragm set OXHD-04
* Probe-filling Electrolyte OXEL-03