METERS
Gas CO
2
/Humidity/Temp. DataLogger
MCH-383SD, CO
2
/Humidity/Temp. Monitor, SD Card Real Time
Datalogger, Patent
Monitor with real time datalogger, save the measuring data along
the time information (year, month, date, hour, minute, second) into
the SD memory card and can be down load to the Excel, extra
software is no need. user can make the further data or graphic
analysis by themselves.
Show CO2 (Carbon dioxide), Humidity and Temperature value in
the same LCD.
NDIR method principal for CO2 (Carbon dioxide) measurement,
available for long term operation. high repeatability and high accuracy.
0.1 %RH resolution for the humidity reading, 0.1 degree resolution
for the Temp. reading.
Used the precision capacitance type humidity sensor, professional
and high accuracy.
SD card capacity: 1 GB to 16 GB.
Large LCD display, easy readout.
Microcomputer circuit, high accuracy.
Low power consumption and long battery life when use battery power.
Power supply from AC/DC 9V adapter in.
RS232/USB computer interface.
Patented
Electrical Specifications (23±5
°
C):
CO2
Range
0 to 4,000ppm
Resolution
1 ppm
Accuracy
± 40 ppm *
1,000 ppm. ± 5% of reading * > 1,000 ppm
3,000 ppm.
± 250 ppm typically * > 3,000 ppm, reference only
Humidity
Range
10 % to 90 % R.H.
Resolution
0.1 % R.H.
Accuracy
70% RH: ± (4% reading + 1% RH). < 70% RH: ± 4% RH.
Temp.
Range
0°C to 50°C, 32°F to 122°F
Resolution
0.1 degree
Accuracy
°C - ± 0.8°C / °F - ± 1.5°F
Model
MCH-383SD
Circuit
Custom one-chip of microprocessor LSI circuit.
Display
LCD size : 60 mm x 50 mm
Measurement
CO2 (Carbon dioxide) Humidity and Temperature
Unit
CO2: ppm, Humidity: %RH, Temp.: °C, °F
CO2 Sensor Structure
NDIR
Temp. Compensation
Automatic temp. compensation for CO2 measurement.
Response Time for CO2 measurement
CO2: < 2 min. typically. * Reach the 63% reading value * Depend the environment
air circulation.
Datalogger Sampling
Time Setting range
5/10/30/60/120/300/600 seconds or Auto. * Default sampling time is 60 seconds. *
The “Auto“ sampling . means when the measuring value is changed (> ± 1 %RH or
> ± 1 °C or ± 50 ppm) will save the data one time only.
Data error no.
0.1 % no. of total saved data typically.
Memory Card
SD memory card. 1 GB to 16 GB.
Advanced setting 1
* SD memory card Format * Set clock time (Year/Month/Date, Hour/Minute/Sec.)
* Set sampling time * Set beeper sound ON/OFF * Set SD card Decimal character
* Select the Temp. unit to °C or °F * Set RS232 data output ON/OFF
Advanced setting 2
* Set CO2 height (meter) compensation value
* Set CO2 height (feet) compensation value
Update Time of Display
Approx. 1 second if measuring data is changed.
Data Output
RS 232/USB PC computer interface. * Connect the optional RS232
cable UPCB-02 will get the RS232 plug * Connect the optional
USB cable USB-01 will get the USB plug.
Power Supply
* Clock backup
* Alkaline or heavy duty DC 1.5 V battery (UM3, AA) x 6 PCs, or equivalent
Power Supply
* Meter & probe
* DC 9V adapter input. (AC/DC power adapter is included)
Operating Temperature
0 to 50°C
Operating Humidity
Less than 90% R.H.
Weight
Meter
240 g/0.53 LB.
CO2 probe
158 g/0.35 LB.
Dimension
Meter: 132 x 80 x 32 mm ( 5.2 x 3.1 x 1.3 inch ) ; CO2 probe: 132 x 38 x 32 mm
Accessories Included
* Instruction manual: 1PC * CO2 probe * AC to DC 9V adapter: 1PC * Main meter
hanging unit (with sticker): 1PC * CO2 hanging unit (with sticker)
Optional Accessories
* SD card (2GB), SD-2GB * USB cable, USB-01. * RS232 cable, UPCB-02.
* Data Acquisition software, SW-U801-WIN.