MRC-2013-Catalog - page 423

423
M
BOD
METERS
Features:
Exchangeable microbe electrode units
make operation smoother and less
time-consuming than loading and
replacing conventional membranes.
Electrode units may be used for one
month continuously or for three
months intermittently with standard
samples. (Electrode units should be
stored in refrigerators when not taking
measurements.)
Front panel allows simple, pushbutton
operation.
Ease of replacement by installing
electrode units or tubing into the
opened door in the front panel.
Anti-reflux valves and antibacterial
tubing reduce the frequency of
maintenance.
Bacteriostatic is mixed into standard
and buffer solutions, so measurements
are stable over the long term.
BODT-1, Biological Oxygen Demand, With Integral Incubator
Model
BODT-1
Measurement method
Biosensor
Measurable
Liquid samples containing soluble organic matter such
as drainage and river water
Measurement range
0-20 mg/l (base Measurement range)
Variable with pump rpm and sample
Contact time
Measurement time
60 min/specimen
(assuming calibration for each specimen)
Calibration
Single-point calibration with glucose glutamic solution
Readout
LCD, 20 characters x 2 lines
Precision up to 0.1 mg/L
Reproducibility
Full-scale within ±5%
Output
Recorder output (0-5 V analog output)
Hot output (0-5 V analog output)
Printer output (Printy 3 serial thermal printer)
Temperature controller
Adjustable within 35°C±0.2°C
Rated power source
AC 100V±10V,50/60HZ,3A max
Dimensions
W370 × D350 ×H300 mm
Weight
14kg
Ambient temperature
5-40°C
Ambient humidity
90% RH max (Without condensation)
Peripherals
Standard solution, buffer solution, antibacterial tubing
set, printer
The Quick BODT-1 takes significant steps to do away with the weaknesses of
biosensor BOD measuring apparatus entailed by trouble and time required
to replace and load microbe membrane and with an exchangeable
electrode. Unit that makes it easy to use right away provides improved
operability and maintainability standard and buffer bacteriostatic solutions
and antibacterial tubing allow stable measurements for long terms The
dedicated shipping and recovery box for Used and replacement electrode
units may be reused any number of times and the exclusion of apparatus
environmentally friendly and easy to use .
BODT-1
Application:
Apply to Surface Water, Sewage, Industrial Wastewater in the BOD.
Features:
Display the trend curve
Printer output
Link to the compute through USB interface
Mercury-free devices to ensure the accuracy & safety of the BOD measurement
Real-time clock display
Automatic collection and storage of data
Depict BOD curves for the user to view
Link to the computer to print curves
No parameter and data loss after sudden cut-off
LCD humanization interface display, easy operation
1-30days of the test time, setting by the keys
Store data and update the result, observe the change rate of BOD in the
LCD screen during any testing time, store 240 data points and 20 groups
data of each sample.
BODT-5, BOD Tester
Model
BODT-5
Measurable
liquid containing soluble organic matter, such as
drainage and river water
Range
0-35, 0-70, 0-350, 0-700mg/L (without dilution)
700-4000 mg/L (need to dilute)
Sample Quantity
6 Bottle Capacity: 500mL
Repeatability
5%
Resolution
0.1mg/L (0-35, 0-70mg/L)
1mg/L (0-350mg/L, 0- 700mg/L)
Dimension
350×400×350mm
Display
LED
Power Supply
AC 220 50Hz
BODT-5
Principle of Operation:
The instrument is designed by the
pressure-difference method, simulating
the degradation of natural organic
matter. 21% of the oxygen in the air on
the top of Test Bottle continuously
replenish the dissolved oxygen
consumed in water, carbon dioxide
(CO2) produced in the process of
organic matter degradation is absorbed
by Sodium hydroxide coming from sealed
cap, pressure sensor monitor constantly
the pressure change of sample bottle.
There is a correlation between
Measurements in regions of low concentration are especially stable.
Changing pump rpm or sample contact time allows measurements in regions of high concentration.
The optional microbe fixation kit enables the use of other microbes and, therefore, a variety of water samples.
Making no use of dangerous reagents and emitting no harmful effluents, the Quick BODT-1 is environmentally friendly.
Biochemical oxygen demand BOD (correspond to the consumption of oxygen in the test bottle) & Gas pressure, such
relevance is dealt with by this instrument, direct display of Biochemical oxygen demand BOD value on the screen.
1...,413,414,415,416,417,418,419,420,421,422 424,425,426,427,428,429,430,431,432,433,...738