PRC30, Multifunction Process Calibrator
PRC30
Features:
Palm-sized double molded housing and large dot-matrix backlit LCD
Measures DC process signals
High accuracy calibration function simulates precision thermocouple
outputs for use in calibrating thermometers, transmitters, controllers or
recorders
Up to fi ve user adjustable calibration presets
Displays output in terms of millivolts or temperature based on
thermocouple tables
24V Loop Power
Zero adjustment, auto zero when power on and change mode
Standard banana I/O ports and mini thermocouple inputs
Large battery bank for extended work cycle
External power adapter for continuous work cycle.
MRC.5.14
Precision source and measure for thermocouples
(8-Types), mA, mV, and V devices
Ordering Information:
PRC30:
ultifunction Process Calibrator
PRC30-NIST:
PRC30 with Certificate of Calibration Traceable to NIST.
Complete with
test leads,
alligator clips,
Thermocouple
calibration
cable with
subminiature
connector,
universal
calibration
Model
Range
Basic Accuracy
DC Source
Current
0 to 24mA, -25% to + 125%
±(0.01% + 1 digit)
Max. Load
1000
@ 24mA
Voltage
0 to 2000mV, 0 to 20V
±(0.01% + 1 digit)
mV/Temp
-5 to 55mV
±(0.01% + 1 digit)
Type J
-50 to 1000°C
±(0.05% + 1°C)
Type K
-50 to 1370°C
±(0.05% + 1°C)
Type T
-120 to 400°C
±(0.05% + 1°C)
Type E
-50 to 750°C
±(0.05% + 1°C)
Type C, R, S
0 to 1750°C
±(0.05% + 1°C)
Type N
-50 to 1300°C
±(0.05% + 1°C)
DC Measure
Current
0 to 50mA, -25% to +230%
±(0.01% + 1 digit)
Voltage
0 to 1999mV/2 to 20V ±(0.01% + 1 digit)/auto range
mV/Temp
-10 to 60mV
±(0.01% + 1 digit)
Type J
-50 to 1000°C
±(0.05% + 1.8°F or 1°C)
Type K
-50 to 1370°C
±(0.05% + 1.8°F or 1°C)
Type T
-120 to 400°C
±(0.05% + 1.8°F or 1°C)
Type E
-50 to 750°C
±(0.05% + 1.8°F or 1°C)
Type C, R, S
0 to 1750°C
±(0.05% + 2°F or 1°C)
Type N
-50 to 1300°C
±(0.05% + 2°F or 1°C)
Loop Power
24V
Meter Dims (LxWxH)
159 x 80 x 44mm
Meter Wt
236g - not including battery weight
CALIBRATION
Applications:
Lab and Field Calibration and
Verification Operations of Sensors,
Thermocouple Probes, Controllers,
Panel meters & Testing tools.
Temperature Control Process
Troubleshooting
Process devices calibration and
simulation: Transducers, Transmitters,
Indicators, Controllers, and Recorders
Field test and service
Bench research and development.
cable with subminiature terminal,
100V-240V AC Universal Adaptor with
4 plugs, six 1.5V AA batteries, and hard
carrying case.