Falcon CALIDUS microGC -012 - page 62-63

Calidus 101-HTResultsCompared to
ConsensusValues
ReportedbyCertificateof Analysis
ValuesShown
Correspond to the cut points
reported in the certificate
Indicate excellent comparison
Calculatedusing raw
chromatograms
LineUpwill improve all values
LineUp use
Absolutely necessary over time for
dataQC automation, no human
can keepupwith~500 runs/day
(~
3minute cycles)
Extendmaintenance interval time
Elevate confidence in the results
Degrees FMeasured Accepted Difference F Limit F
IBP
240
239
1.0
13.7
5
304
304
0.0
6.8
10
349
349
0.0
7.4
15
395
393
2.0
8.1
20
437
435
2.0
8.6
25
472
469
3.0
8.5
30
500
499
1.0
8.5
35
528
526
2.0
8.1
40
554
552
2.0
7.7
45
578
576
2.0
7.7
50
595
594
1.0
7.7
55
611
610
1.0
7.7
60
629
629
0.0
7.7
65
649
649
0.0
7.7
70
669
669
0.0
7.7
75
690
690
0.0
7.7
80
713
712
1.0
7.7
85
737
736
1.0
7.7
90
765
764
1.0
7.7
95
805
803
2.0
9.0
FBP
887
887
0.0
21.2
1...,42-43,44-45,46-47,48-49,50-51,52-53,54-55,56-57,58-59,60-61 64-65,66-67,68-69,70-71,72-73,74-75,76-77,78-79,80-81,82-83,...