CALIDUS
tm
201microGasChromatograph
GCanalysis for virtuallyany fixedgas andhydrocarbonsup toC
44
for laboratory, at-line,
transportableor onlineuse
CALIDUSModel 201 -
comprisedof 4modules
A singleSampleProcessingUnitwitha standard split/splitless injectionport (1:1up to1:200) suitablegasand liquid
samples viaeither syringe through the septum injections, optional gas, liquidor headspaceauto-sampler, or
automated samplingvalves. The inlet includes septumpurge toprevent bleed components fromentering the
system.
TwoProgrammedTemperatureColumnModules (PTCM) in series containing the resistivelyheated steel capillary
chromatography columnwithnecessaryhardware, softwareandelectronic control toenable temperature
programming from0.1
o
C to5
o
Cper second from5
o
Caboveambient to350
o
Cdependingon themaximum
temperature capabilityof the columnmaterial selected. Each columnmodule is independently controlledby the
methodand canbeanyof theavailable column types.
A singleDetectorModule incorporatingeither amicro Flame IonizationDetector (FID) ormicroThermal Conductivity
Detector (TCD)with thenecessaryhardware, softwareandelectronic control toprovidedetector temperature
control, digital output signal andadditionally forproper FID fuel supplypressureandauto-ignition.
ThemicroFID is a fullydigital carbon/hydrogenbonddetectorusing thehydrogen flame toburn the sample
components. Ituses anelectrometer to sense the current changes in the flame cell due to chromatographic
component elution. Thedata rate is100Hz.
ThemicroTCD is a fullydigital, universal detector. TheTCD consistsof a constant temperature filament that
senses change inpower required tohold the filament temperature constantwhen chromatographic
components elute. Thepowermeasurement isused todetermine theamount of the component eluting
from the column. Thedata rate is50Hz.
CALIDUS is controlledwithChromPerfect chromatographydata system fully integratedwith LineUp runningona
WindowsPC.
100AEI Drive
Fairlea,WV24901
P.O. Box518
Ronceverte,WV 24971
Phone: 304-647-5860
Fax: 304-645-4006
E-mail:
SIGNIFICANCEANDUSE
TheModel 201provides a simpleultra-fast analysis (10 to50 times faster than conventional lab
or processGC’s) of fixedgases andhydrocarbons toC
44
. Useof twodifferent PTCMs in series,
for exampleonepolar andonenon-polar columnmaterial enables leveraging the selectivity
differences for enhanced separations. Secondly, using two identical PTCMs virtuallydoubles
the column length.
Theanalyses areused for product specifications testing, product safety, environmental
measurements, process control, catalyst protection, educational tools, spot checksof fuels and
manymore.
Faster, Smaller, Smarter, Easier,Greener