RESPONSE TIME

NOISE

The LD8000 has a fast response time to ensure a good

purity control in the different applications of such analysis.

However, such good result is always dependant of the

installation along with the instrument. The stream selector

system is critical to be able to achieve such good response

time. We strongly suggest to use the LDGSS from LDetek

which has been designed specifically for such application.

The electronic for the acquisition of the detector has been

designed to avoid any noise that could influence the reading

of the analyzer. But more importantly, the design of the

detector itself is very important to minimize the noise as much

as possible. The position of the optical filter along with the

photo-diode is critical to get the specification of the LD8000.

Bad response of those components means electrical gain

increase to achieve the measurement desired and at the

same time the noise level will be amplified. When mounting

the detector, special care is done by LDetek specialist.

The amplifier board used to manage the signal from the

photo-diode was also particularly well designed. This board

takes very low voltage to amplify it on a scale up to 0-10 volts.

Analyzer Performance

The design of the LD8000 brings outstanding performances for the market demand. Even if this analyzer is used for ppm

measurement, we get stability, accuracy and noise level for ppb analysis. With the design described above, some tests have been

done to show the performances in different situations. Those performances have been tested with argon sample, since it is the most

popular application on the market.

Figure 10

:

*LD8000 Reponse time

Figure 11

:

*Noise level acquired on the analog output

with a sample gas of about 0.104 ppm

Note:

This result has been obtained in LDetek facilities with the LDGSS stream selector device with a well purged system according to LDetek standard methods.

*