![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0038.jpg)
4
/
8
©Copyright 2012 LDetek inc. • 271, Saint-Alphonse Sud, Thetford Mines (Quebec) Canada G6G 3V7
Flow Transducer
The flow transducer is the biggest source of contamination for
such analyzer. To avoid any problem with such device, the
transducer is installed at the end of the flow path after the
detector and is connected to the microcontroller unit to give
information to the micro-valve to control a stable flow.
So why not using a MFC after the detector? That would avoid
any chance of contamination or leakage? Actually, the PED
must work at atmospheric pressure and it must not be
pressurized. Since the PED is a cell made of quartz, it can
break with a backpressure coming from the vent.
Furthermore, you want to keep a constant pressure in the
detector to get a stable plasma.
NO NEED OF ZERO CALIBRATION GAS!
An option is available that avoids the need of a pure reference gas to make the zero calibration. The LD8000 can be designed in a way
that components are added inside the analyzer to generate zero gas.
Only the sample gas is connected to the instrument and zero calibration can be achieved. A valve’s system is used to change the flow
path to the detector. During zero calibration, the sample goes to a small gas purifier to get pure argon or helium.
There is no consumable in such configuration. Since the integrated gas purifier is used only during zero calibration and the valve
system isolates it, the lifetime of the purifier is extended. Such system will save the needs of pure argon supply, gas regulator, piping,
external purifier, bypass valve for the purifier, etc. The payback is quite evident!
Figure 7
:
Flow path of the LD8000
Figure 5
:
Standard valve performance
Figure 6
:
LDetek valve performance
Sample vent
1/8 compression
fittings
Flow Transducer
Plasma Detector
+
Humidity Control
Device
Micro-Valve
Sample inlet
1/8 compression
fittings