PRINTED IN CANADA

DIMENSIONS:

TOP VIEW

482 mm (18.98 in)

13 mm

(0.51 in)

446 mm (17.56 in)

595 mm (23.43 in)

FRONT VIEW

482 mm (18.98 in)

101 mm

(4 in)

177 mm

(7 in)

CUT OUT DIMENSIONS

FOR RACK MOUNTING

450 mm (17.72 in)

101 mm

(4 in)

O 8 mm (0.31 in.)

183 mm

(7.2 in)

SIDE VIEW

595 mm (23,43 in)

177 mm

(7 in)

(MOST POPULAR APPLICATIONS)

H2

Ar

O2

N2

CH4

CO

CO2

NMHC

N2O

C2H2

C2H4

C2H6

C3H8

Ne

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

H2

Ar

O2

N2

He

CO2

NH3

CAPABILITIES:

*some application requires more than one chassis/Multidetek

Other measurement possible, please consult factory

Samples

Impurities

DETECTOR TYPE

Plasma Emission Detector design (PlasmaDetek)

RANGE

Application dependant

CARRIER

Argon and Helium

REPEATABILITY

< 1% full scale

ACCURACY

Better than ±1% full scale

STANDARD FEATURES

• Manual or autoranging (user selectable)

• Microprocessor controlled

• 5.6” TFT intelligent LCD module with Touch Screen

• Self diagnosis system with auto-resolve alarm

• 4-20 mA isolated output

• Alarm Historic

• Digital system status ouput for

remote monitoring (dry relay contact)

OPTIONS

Serial port: RS-232 / 422 / 485 / Profibus for monitoring

SAMPLE PRESSURE REQUIREMENTS

10 to 30 PSIG

CARRIER PRESSURE REQUIREMENTS

100 PSIG

OPERATING TEMPERATURE

10

°

C to 45

°

C

SUPPLY

POWER CONSUMPTION

DRIFT

115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz

200 Watts (application dependant)

< ± 1% over 24 hours

GAS CONNECTIONS

Sample: 1/8’’ compression fittings or 1/8 VCR - Vent: 1/8’’ compression fitting

CALIBRATION GAS

Zero: LDP1000 purified gas (Getter) - Span: 70% to 90% of the scale

SPECIFICATIONS: