The selectivity mode simplifies the chromatography

and can be configured for specific application.

By controlling the temperature of the doping agent injection device,

better stability of the measurement is achieved. The temperature is

controlled by the PlasmaDetek controller and can be adjusted for

the specific application. The use of different doping agents is

application dependant is part of the selectivity mode.

SELECTIVE AND NON-SELECTIVE

CONFIGURATION

DOPING AGENT INJECTOR TEMPERATURE

CONTROLLED

2

The PlasmaDetek design avoids the creation of phantom

peaks that occurs in conventional ionization detector. Its

unique design made of a monolitic quartz makes the detector

dead volume free. It is not affected by pressure or flow swings.

PED DESIGN

WITH NO DEAD VOLUME

Conventional ionization detector

100ppm O

2

& 10ppm N

2

in He

O

2

O

2

O

2

N

2

N

2

N

2

12ppm N

2

in balance Oxygen

Plasmadetek

100ppm O

2

& 10ppm N

2

in He

12ppm N

2

in balance Oxygen

CARRIER

GAS

LIGHT EMISSION

GENERATOR

HIGH INTENSITY

HIGH FREQUENCY