Each piston will be supplied with a porous frit, which can be removed for cleaning or change out. It is important to note that
there is a specific orientation to the frit. See diagram below.
To remove existing frit, insert frit ejector into inlet end of piston and push frit out gently. If the frit does not dislodge easily, put
piston assembly in hot water (maximum of 120°C) to expand plastic and allow for easier frit removal.
Frit Ejection Method
Frit Installation