Table of Contents Table of Contents
Previous Page  401 / 413 Next Page
Information
Show Menu
Previous Page 401 / 413 Next Page
Page Background

Chemical/Petrochemical

GC Analysis of Total Reduced Sulfurs at ppbv Levels

Using an

Rxi™-l

ms Column and Sulfur Chemiluminescence Detection

by Silvia Martinez, In novations Chemist

• Reliable results for ppbv concentrations of highly active sulfur compounds.

Inert, low bleed column resolves all analytes.

• Column compatible with SCD and other sulfu r-specific detectors.

Through the Clean Air Act, the United States Environmental Prot ection Agency (US EPA) regulates and lim its the emission of toxic

air pollutants. The determinat ion of total redu ced sulfurs, as requ ired by CFR Title 40, requi res the use of methods and equipment

capable of providing full resolution as well as high sensitivity. Method s TO- IS, TO -I 6 and TO-I 6A describe GC proced ures that

apply to th e determination of redu ced sulfurs from stat iona ry sources, such as recovery furn aces, lime kilns, smelt dissolving tanks,

fuel gas combustion devices, tail gas contro l units, and others. Method TO - I6 specifies detectable conc entrations of ppbv levels for

dimethyl disulfide, dimethyl sulfide, hydrogen sulfid e, and me thyl merca pt an. Wh ile th ese meth ods do not specify th e ana lytical GC

column to use, they do state that the column must resolve the sulfur compoun ds.

Our new 100% dimethylpolysiloxane column, th e

Rxi' ">Ims column, provides th e ult ra-low bleed

required for low level detection and quant ification

of sulfur compounds. Its exceptional inertne ss

allows complete sepa ratio n of th ese very reactive

compoun ds, with excellent peak shape, at ppbv

levels. Wh en thi s colum n is coupled wit h a sulfur

chemiluminescence detector (SCD), th e analysis is

fast and simple.

For our examp le ana lysis, we collected a 20mL

sample of a gaseo us mixture of hydrogen sulfide,

carbo nyl sulfide, met hyl me rcaptan, ethyl mercap­

tan, and dimeth yl sulfide in helium, using a

Sulfinertw-treated stainless steel sample loop. We

transferred the sample to a SilcoCa n" air moni ­

toring caniste r and press ur ized the can to 30psig

with d ry nitrogen. The

Sulfinertw

passivation

treatment on both the sample loop and canister

prevent ed adsorption losses of the highly active

sulfur compo unds. We intro duced a ImL aliquot

of the diluted gaseous mixture int o th e Rxi™-I ms

column via a second

Sulfinertv-treated

stainless­

steel sample loop, using helium as a carr ier, and

analyzed the sample isotherma lly at 30°C.

Figure I shows the chromatography for the

red uced sulfu r compounds, demon strating full

resolution in less than S minutes. For collecting,

storing, and analyzing active sulfur compounds at

levels as low as single parts per billio n, the per­

forma nces of

Sulfinertw

passivated containers and

transfer systems, and inert RxiTM_ Ims columns,

simply can't be equaled.

Rxi™-1ms Columns (fused silical

(Crossbond 100%dimethyl polysiloxane)

ID

df

Ipm)

temp.limits

length

cat

#

price

0.32mm 4.00 -60 to

330/350'C

30-Meter 13396

$460

2006.03

Figure 1 Total resolution of reduced sulfur compounds, in lessthan

5 minutes, using an Rxi™-l ms column .

1. hydrogensulfide

2. carbonylsulfide

3. methyl mercaptan

4. ethyl mercaptan

5. dimethyl sulfide

Iii" iii iii i i" " i' i I i i ' i' " ii i'

i '

i " ii' I ' i i " ii ' ' I

U

W

~

~

~

Time(min)

GC AR00861

Column:

Rxi™ -lms, 30m, 0.32mm 10,

4.00/lm

(cat.# 13396)

Sample:

hydrogensulfide, carbonyl sulfide, methyl mercaptan,ethyl mercaptan,

dimethylsulfide,100ppbveachin helium

Inj.:

1.mLsplitless, direct

Sample looptemp.: 30' C

Carrier gas:

helium, constant pressure

Linear velocity:

48cm/sec.

@

30'C

Oven temp.:

30'C

Del.:

sulfur chemiluminescence detector

Del.temp.:

800'C

Sample storage

&

transfer:

SilcoCan' air monitoringcanister withSiltek

O

treated1/4" valve(cat.# 24182-650); Sulfinert"

treatedgas sample loop, l ee(cat.# 22848); Sulfiner!" treatedgas sample loop, lOcc (custom

order)

for

more

info

SilcoCan™Canisters

Thebest alternative for ambient air monitoring

Recovery data for low ppb levels of active sulfur-containing compounds

show why

Silcocan"

canisters are the best choice for monitoring

TO-14,TO-IS, or reactive sulfur compounds. lit cat# S90llA

800-356-1688 °

www.restek.com

• 20 •

Website :

www.chromtech.net.au

E-mail :

info@chromatech.net.au

TelNo : 03 9762 2034 . . . in AUSTRALIA