In Agilent split/splitless injection ports, it can be

difficult to make and maintain a good seal with a

conventional metal inlet disk. The metal-to-metal

seal dictates that you apply considerable torque to

the reducing nut, and, based on our testing, this

does not ensure a leak-tight seal. Over the course of

oven temperature cycling, metal seals are prone to

leaks, which ultimately can degrade the capillary

column and cause other analytical difficulties.

Our patented Dual Vespel® Ring Inlet Seal greatly

improves injection port performance—it stays sealed, even after repeated tem-

perature cycles, without retightening the reducing nut! This seal features two

soft Vespel® rings, one embedded in its top surface and the other embedded in

its bottom surface. These rings eliminate the need for a washer, and ensure

very little torque is needed to make a leak-tight seal. The rings will not harm

the critical seal in the injector body, or any other surface, and are outside the

sample flow path. Tests using a high sensitivity helium leak detector show Dual

Vespel® Ring Inlet Seals will seal equally effectively at torques from 5 in. lb. to

60 in. lb. (Figure 1).

Why trust a metal-to-metal seal when you can make leak-tight seals quickly

and easily—and more reliably—without a washer, with a Restek Dual Vespel® Ring Inlet Seal. Use a stainless steel seal for analyses of

unreactive compounds. To reduce breakdown and adsorption of active compounds, use a gold-plated or Siltek®-treated seal. The

gold surface offers better inertness than untreated stainless steel; Siltek® treatment provides inertness similar to that of a fused silica

capillary column.

Figure 1

The Dual Vespel® Ring Inlet Seal

achieves leak-tight seals even at low

torque, reducing the chance of leak-related

problems.

22

2007 vol. 4

GC Accessories

Dual Vespel® Ring Inlet Seals

Washerless, Leak-Tight Seal for Agilent GCs

• Prevents oxygen from permeating the carrier gas, increasing column lifetime.

• Vespel® ring in top surface reduces operator variability by requiring minimal torque to seal.

• Vespel® ring in bottom surface simplifies installation—eliminates the washer.

Washerless, leak-tight seals for Agilent GCs

0.8mm ID Dual Vespel

®

Ring Inlet Seal

2-pk.

/price

10-pk.

/price

Gold-Plated

21240

$58

21241

$264

Siltek

®

Treated

21242

$58

21243

$264

Stainless Steel

21238

$45

21239

$195

1.2mm ID Dual Vespel

®

Ring Inlet Seal

2-pk.

/price

10-pk.

/price

Gold-Plated

21246

$58

21247

$264

Siltek

®

Treated

21248

$58

21249

$264

Stainless Steel

21244

$45

21245

$195

Dual Vespel® Ring Cross-Disk Inlet Seals for Agilent GCs

• Ideal for high-flow split applications.

• Washerless, leak-tight seals.

0.8mm ID Dual Vespel

®

Ring Cross-Disk Inlet Seal

2-pk.

/price

10-pk.

/price

Gold-Plated

22083

$66

22084

$295

Siltek

®

Treated

22085

$66

22086

$295

Stainless Steel

22087

$53

22088

$239

Leak Rate (Log

10

atm cc/sec.)

0

20

10

30

40

0.01

1E-3

1E-4

1E-5

1E-6

1E-7

1E-8

1E-9

1E-10

50

60

Dual Vespel® Ring Inlet Seal

stainless steel

original equipment

inlet seal

Torque (in. lbs.)

Dual Vespel® Ring Inlet Seals are available in gold plating, stainless steel, and Siltek® treated.

Vespel® ring seal on top

and bottom surfaces!

Patented.

new

!