Table of Contents Table of Contents
Previous Page  196 / 413 Next Page
Information
Show Menu
Previous Page 196 / 413 Next Page
Page Background

• 22 •

Tech Tip

Under Pressure?

Reduce System Stress by Backflushing Your HPLC Column

By Tim Herring, Technical Service

Experiencing a higher pump pressure than usual? Or perhaps a complete pressure shut-down of the system has occurred, even after

replacing the in-line frit and guard column. High pump pressures can be caused by heavily retained impurities building up within

the head of the analytical column. Such contamination can cause poor chromatography, usually in the form of broad, split, or mis-

shapen peaks, and ultimately can compromise results. Backflushing a contaminated analytical column using the following procedure

can help restore column performance and reduce pump pressure and system strain.

If back pressure is abnormally high, first take the column out of the equation by disconnecting it from the system altogether. Install

a union and run the pumps to verify that the back pressure problem is due to the column, and not to the HPLC system. If the pres-

sure is normal, then the column is most likely the cause of the high back pressure. To address this, reverse the column flow and rinse

(backflush) the column to remove the contaminants from the inlet frit and column head. This will move the contaminants down the

path of least resistance, instead of forcing them further into the analytical column. Reverse rinse into a waste beaker at low flow (e.g.

0.5mL/min. for a 4.6mm ID column) for 10-15 minutes initially, and then increase the flow to 1.5-2 times the optimal flow (1.5 to

2.0mL/min. for a 4.6mm ID column). Do not reconnect to the detector when backflushing the column. Rather, flush the waste

stream into a beaker so that the detector cell is not contaminated by impurities or obstructed by particulate build-up.

Solubility is a key issue when backflushing columns, so remember the old adage, “like

dissolves like”. For example, if the contaminants are suspected to be oily or

hydrophobic in nature, then backflush with a strong, nonpolar solvent such as

hexane. If the contamination is polar (a salt for instance), then use a polar

solvent, such as water or methanol. Solvent miscibility also needs to be

considered, so be sure to use solvents that are miscible with one anoth-

er. If in doubt, use isopropanol (IPA) as an intermediary solvent

between solvent wash steps, as it is miscible with all common sol-

vents. This is particularly true when switching from typical nor-

mal phase solvents (such as hexane) to reverse phase solvents

(such methanol, acetonitrile, or water) and vice versa. Note that

10 to15 column volumes are generally necessary at each step to

remove all traces of immiscible solvents prior to the next step.

If the contaminants are unknown, or vary in chemistry, a series

of solvent washes will provide an array of differing chemical

interactions and maximize the removal most types of contami-

nation. The solvent order presented in Table I considers miscibil-

ity, polarity, and eluotropic strength and is a very effective series

for removing most contaminants. Column backflushing, with

proper solvent selection, is a simple way to regenerate analytical

columns, improving column performance and reducing system stress.

Contact Restek Technical Service

at

support@restek.com

or 800-356-1688

with questions on backflushing, or any other

technical area. At Restek, we are here to help you!

Reversed phase series:

Normal phase:

A. 1% glacial acetic acid in methanol and water (50:50)

A. isopropanol

B. methanol

C. chloroform

D. hexane (or heptane)

E. methylene chloride (dichloromethane)

F. methanol

Table I

Restore column performance by backflushing with

recommended solvent washes.