Table of Contents Table of Contents
Previous Page  63 / 298 Next Page
Information
Show Menu
Previous Page 63 / 298 Next Page
Page Background

• 13 •

800-356-1688 •

www.restek.com

2005 vol. 1

New GC Column for PCB Congeners or

Aroclor® Mixes: Rtx®-PCB

Exclusive Polymer with Unique Selectivity

by Gary Stidsen, GC Columns Marketing Manager

•Unique polymer for PCBs analysis by GC/ECD or GC/MS.

•Good results for other semivolatiles.

• Low polarity and inertness for active compounds.

• Thermally stable to 340ºC.

Rtx

®

-PCB columns contain a proprietary polymer

that has provided unique separations for PCB

congeners, and can be used with electron cap-

ture detection or mass spectrometry. Figure 1

shows the excellent peak shape obtained for the

PCB congeners in several Aroclor

®

mixes. In our

initial review of data for these columns, we dis-

covered that by using an Rtx

®

-PCB column in a

GC/MS analysis, European PCB congeners can

be analyzed without interference from other

congeners. Table 1 is a shortened list of the PCB

congeners, showing those that elute near the

European PCB congeners, which are indicated

by an “x”.

Figure 1

Excellent separation and peak

shape for PCBs in three Aroclor® mixes, using an

Rtx®-PCB column.

Column:

Rtx

®

-PCB 30m, 0.25mm ID, 0.25µm (cat.# 13223)

Sample:

200ng/mL Aroclor

®

1242 (cat.# 32009); Aroclor

®

1254 (cat.# 32011); Aroclor

®

1260 (cat.# 32012)

Inj.:

1.0µL splitless (hold 0.75 min.), 3.5mm ID single

gooseneck inlet liner (cat.# 20962)

Inj. temp.:

250ºC

Carrier gas:

hydrogen, constant pressure

Linear velocity:

71cm/sec. @ 110ºC

Oven temp.:

100ºC (hold 1.0 min.) to 300ºC @ 10ºC/min. (hold 4

min.)

Det.:

ECD @ 310ºC

Rtx

®

-PCB Columns (fused silica)

ID

df (µm)

temp. limits*

20-Meter

30-Meter

60-Meter

0.18mm 0.18

30°C to 320/340°C 41302

$445

41304

$470

0.25mm

0.25

30°C to 320/340°C

13223

$445

13226

$470

0.32mm 0.50

30°C to 320/340°C

13239

$525

unique polymer

technology

new

!

Aroclor

®

1242 on Rtx

®

-PCB

Aroclor

®

1254 on

Rtx

®

-PCB

Aroclor

®

1260 on

Rtx

®

-PCB

Table I

Coelutions do not interfere with analysis of European PCB congeners (“x”) on Rtx®-PCB:

only congeners not measurable by MS are indicated in boxes.

Eur

IUPAC #

Cl #

T

R

(min.)

Resolution

Assessment

53

4

14.11

31

3

14.14

0.5

x

28

3

14.23

1.4

Measurable by MS

33

3

14.27

0.6

51

4

14.29

0.4

20

3

14.30

0.1

45

4

14.54

3.8

46

4

14.71

43

4

14.88

2.7

x

52

4

14.94

1.0

Measurable by MS

48

4

15.01

1.2

49

4

15.08

1.0

89

5

17.29

84

5

17.30

0.2

56

4

17.34

0.6

x

101

5

17.35

0.2

Measurable by MS

99

5

17.50

2.3

60

4

17.52

0.4

123

5

19.18

109

5

19.19

0.2

134

6

19.24

0.7

133

6

19.28

0.7

x

118

5

19.35

1.0

Measurable by MS

131

6

19.35

0.0

146

6

19.47

1.9

122

5

19.53

1.0

114

5

19.65

1.8

x

153

6

19.66

0.2

132

6

19.77

1.7

179

7

19.88

1.8

130

6

20.31

164

6

20.33

0.3

178

7

20.45

1.9

x

138

6

20.47

0.3

Measurable by MS

163

6

20.51

0.6

129

6

20.56

0.7

158

6

20.60

0.6

172

7

21.99

156

6

22.07

1.3

157

6

22.18

1.6

x

180

7

22.19

0.3

Major congener 180 should

193

7

22.23

0.6

not be terribly biased by 193.

200

8

22.30

1.1

191

7

22.37

1.1

Mix of Aroclor

®

1242-1254-1262 used for resolution check.

Relaxed resolution criteria based on visual inspection of closely eluting congeners.

Rtx

®

-PCB is the newest member of a family of

new polymer phases that undergo rigorous qual-

ity assurance measures to ensure every column

meets exacting standards and that performance

is reproduced from column to column. Specified

column parameters include film thickness, bleed

(at 320ºC), inertness, plates per meter, and

retention time indices. These measures assure

you of the highest quality columns available.