GC_EV00699

GC_EV00698

Rtx-5Sil MS columns

are listed on page 9!

• 2 •

www.restekcorp.com

RESTEK Advantage

800-356-1688

response are excellent for the very active compounds endrin, DDT, and

methoxychlor, peaks 89, 95, and 102, respectively.

Low-bleed Rtx

®

-5Sil MS columns ensure low detection limits and excellent instru-

ment stability in semivolatiles analysis.

We incorporate 106 Method 525.2 target compounds in six new, stable mixtures.

The concentration of each component in these mixes, 1000µg/mL,

1

is consider-

ably higher than in mixes from other sources, and significantly more analyses can

be conducted from each ampul of material. We have formulated the new calibra-

tion mixes by chemical class: semivolatiles, PCB congeners, organochlorine pesti-

cides, nitrogen/phosphorous pesticides.

2

Pentachlorophenol is included in the

semivolatiles mix at a concentration four times higher than the other analytes, as

required by the method. For the convenience of our customers we include hep-

tachlor epoxide isomer A, an analyte not on the target list for Method 525.2, in

Chlorinated Pesticides Mix #2. Six of the nitrogen/phosphorous pesticide analytes

on the target list are unstable in aqueous matrices; we combine these analytes in

a separate mix (cat.# 32423). We do not include two compounds, disulfoton sul-

foxide and disulfoton sulfone, in either our new mixes or our existing mixes. In

water, disulfoton is rapidly oxidized by chemical reaction to the analogous sulfox-

ide and sulfone. Consequently, the sulfoxide and sulfone cannot be included in the

mix with disulfoton. We offer individual solutions of the two analytes — please

refer to our catalog or web site.

3

In addition to calibration mixes for Method 525, we also have all needed quality

control standards: internal and surrogate standards, a fortification recovery stan-

dard, and a GC/MS performance check standard. New Method 525.2 Surrogate

Standards Mix includes optional pyrene-d10. All new mixes are described on page 3.

References

1

Components in Method 525.2 PCB Congeners Mix (cat.# 32420) are at 200µg/mL each.

2

Previously available mixes for Method 525 include Organochlorine Pesticides Mix AB #3

(cat.# 32415), individual Aroclor

®

PCB solutions (cat.# 32075, 32077, 32079, 32081, 32083, 32085,

32087), and TCLP Toxaphene (cat.# 32015).

3

Catalog # MET 652-A and MET 652-B, respectively.

Figure 1

- Low bleed and excellent

resolution of complex semivolatiles

sample, using an Rtx

®

-5Sil MS column.

1. isophorone

2. 2-nitro-

m

-xylene

3. dichlorvos

4. hexachlorocyclopentadiene

5. EPTC

6. mevinphos

7. butylate

8. vernolate

9. dimethylphthalate

10. etridiazole (Terrazole

®

)

11. 2,6-dinitrotoluene

12. pebulate

13. acenaphthylene

14. acenaphthylene-d10

15. chloroneb

16. 2-chlorobiphenyl (BZ#1)

17. tebuthiuron

18. 2,4-dinitrotoluene

19. molinate

20. diethyl phthalate

21. fluorene

22. propachlor

23. ethoprop

24. cycloate

25. trifluralin

26. chlorpropham

27.

α

-BHC

28. 2,3-dichlorobiphenyl (BZ#5)

29. hexachlorobenzene

30. atraton

31. prometon

32. simazine

33. atrazine

34. propazine

35.

β

-BHC

36. pentachlorophenol

37.

γ

-BHC (lindane)

38. terbufos

39. diazinon

40. pronamide

41. phenanthrene-d10

42. chlorothalonil

43. phenanthrene

44. methyl parathion OA

45. disulfoton

46. terbacil

47. anthracene

48.

δ

-BHC

49. 2,4,5-trichlorobipenyl

(BZ#29)

50. metribuzin

51. alachlor

52. simetryn

53. ametryn

54. prometryne

55. heptachlor

56. terbutryn

57. di-

n

-butylphthalate

58. bromacil

59. 2,2',4,4'-tetrachloro-

biphenyl (BZ#47)

60. metolachlor

61. chlorpyrifos

62. cyanazine (Bladex)

63. DCPA methyl ester

(Dacthal

®

)

64. aldrin

65. triadimefon

66. diphenamid

67. MGK-264

68. merphos

69. heptachlor epoxide (isomer B)

70. 2,2',3',4,6-pentachloro-

biphenyl (BZ#98)

71. heptachlor epoxide (isomer A)

72. stirofos (tetrachlorvinphos)

73. butachlor

74.

α

-chlordane

75. fenamiphos

76. pyrene-d10

77.

γ

-chlordane

78. endosulfan I

79. pyrene

80.

trans

-nonachlor

81. napropamide

82. 4,4'-DDE

83. tricyclazole

84.

p

-terphenyl-d14

85. 2,2',4,4',5,6'-hexachloro-

biphenyl (BZ#154)

86. carboxin

87. dieldrin

88. chlorbenzilate

89. endrin

90. 4,4'-DDD

91. endosulfan II

92. endrin aldehyde

93. norflurazon

94. butyl benzyl phthalate

95. 4,4'-DDT

96. bis(2-ethylhexyl)adipate

97. endosulfan sulfate

98. hexazinone

99. triphenylphosphate

100. endrin ketone

101. 2,2',3,3',4,4',6-hepta-

chlorobiphenyl (BZ#171)

102. methoxychlor

103. 2,2',3,3',4,5',6,6'-octa-

chlorobiphenyl (BZ#207)

104. benzo(a)anthracene

105. chrysene-d12

106. chrysene

107. bis(2-ethylhexyl)phthalate

108. fenarimol

109.

cis

-permethrin

110.

trans

-permethrin

111. benzo(b)fluoranthene

112. benzo(k)fluoranthene

113. fluridone (Sonar

®

)

114. benzo(a)pyrene

115. perylene-d12

116. dibenz(a,h)anthracene

117. indeno(1,2,3-cd)pyrene

118. benzo(ghi)perylene

Rtx

®

-5Sil MS 30m, 0.25mm ID, 0.25µm (cat.# 12723)

Sample:

US EPA Method 525.2 Mix

1µL 10ppm (20ppm IS)

Standards used: cat. #s 31824, 32420,

32421, 32422, 32423, 31825, 31826,

31828, 32291

Inj.:

1.0µL 30psi 0.4min. pulsed splitless

(hold 0.4 min.),

4mm Drilled Uniliner

®

(cat.# 21055)

Inj. Temp.:

300°C

Carrier Gas:

helium, constant flow

Flow Rate:

0.8 mL/min.

Oven Temp.:

35°C (hold 1 min.) to 270°C @

18°C/min. (hold 0 min.) to 330°C

@ 5°C/min. (hold 1 min.)

Det:

Agilent 5973 GC/MS

Transfer Line Temp.: 280°C

Scan Range:

40–550 amu

Solvent Delay:

4.0 min.

Tune:

DFTPP

Ionization:

EI