Figure 1
The MXT
®
-1HT Sim Dist column exhibits
low bleed, which gives accurate
quantitation at high temperatures.
Innovators of High Resolution
Chromatography Products
the
in this issue
Rt-XLSulfur
™
GC Column and
Sulfinert
™
System for Trace Sulfur in
Beer Analyses
…pg. 6–7
Stx
®
-CLPesticides GC Column Pairs for
Improved Endrin Response
…pg. 8–9
Stabilwax
®
-DA GC Column for Analysis of
Flavor Components in Whiskey
…pg. 10-11
Ethanol Analytical Reference Materials
for Blood Alcohol Analyses
…pg. 11
UST Analytical Reference Materials
…pg. 12
Natural and Refinery Gas Analytical
Reference Materials
…pg. 13
New InfraRed
™
Septa
…pg. 14
Cool Tools
…pg. 15
Behind the Scenes
…pg. 16
800-356-1688
814-353-1300
www.restekcorp.comFall
2001
by Neil Mosesman, GC Columns Product Marketing Manager,
and Dinesh V. Patwardhan, Ph.D., Senior Research Chemist
Low-Bleed, High-Temperature
Column for ASTM Method D-6352
MXT
®
-1HT Sim Dist GC Column
Advantage
MXT
®
-1HT Sim Dist
GC Column for High-
Temperature Analyses
…pg. 1–2
Pinnacle
II
™
C18
and Cyano HPLC Columns
for Trace-Level Explosives
Analyses
…pg. 3–4
Rtx
®
-5Sil MS Columns for
Fast Semivolatiles Analyses
…pg. 5
The American Society for
Testing and Materials
(ASTM) is an organiza-
tion that publishes con-
sensus standards for
materials, products, and
services. Because ASTM
methods are developed
by consensus from the
laboratories following the
method, they are recog-
nized worldwide. ASTM
Method D-6352 is a gas chromatography (GC)
method developed for the determination of petrole-
um distillates with a boiling point range of 174°C to
700°C. Often referred to as high-temperature simu-
lated distillation or Sim Dist, this method requires a
capillary column capable of withstanding high GC
oven temperatures, up to 430°C. This presents
many challenges for analysts because most capillary
✔
Durable metal tubing and high-temperature polymer
allows analysis of hydrocarbons up to C110.
✔
Low bleed at high temperatures help achieve accurate quantitation.
✔
Meets all criteria of ASTM Method D-6352.
new
10
20
30
40
min.
0
0
10
20
30
40
50
60
70
pA
5m, 0.53mm ID, 0.10µm MXT
®
-1HT (cat.# 70100)
GC:
Agilent 6890;
Injector:
on-column (track oven);
Carier gas:
helium (constant pressure);
Head pressure:
1.0psi;
Linear velocity:
60cm/sec.;
Flow rate:
7.8mL/min.;
Detector:
FID @ 430°C;
Make-up gas flow:
40cc/min.;
Temp. program:
40°C
to 430°C @ 10°C/min. (hold 30 min.)
430°C
Figure 2
The MXT
®
-1HT Sim Dist column allows analysis of hydrocarbons up to C110
in less than one hour.
20
30
40
50
min.
C30
C40
C44
C50
C60
C70
C80
C90
C100
C110
5m, 0.53mm ID, 0.10µm MXT
®
-1HT Sim Dist
(cat.# 70100)
Sample:
1.0µL Polywax 1000 standard
(cat.# 36227);
Solvent:
carbon disulfide;
GC:
Agilent 6890;
Injector:
on-column (track
oven);
Carier gas:
helium (constant pressure);
Head pressure:
1.0psi;
Linear velocity:
60cm/sec.;
Flow rate:
7.8mL/min.;
Detector:
FID
@ 430°C;
Make-up gas flow:
40cc/min.;
Temp. program:
40°C to 430°C @ 10°C/min.
(hold 30 min.)
GC_PC00530