Volatile Organics by EPA 8260B on Rtx®-VMS

GC COLUMNS |

APPLICATION-SPECIFIC COLUMNS

Environmental

Rtx®-VMS Columns

(fused silica)

(proprietary Crossbond® phase)

• Application-specific columns for analyzing volatile organic pollutants by GC-MS

including Methods TO-15, TMS, and EPA 8260.

• Complete separation of U.S. EPA Method 8260 compounds in less than 10 minutes.

• Stable to 260 °C.

• No known equivalent phases.

Rtx®-VMS columns offer lower bleed, better selectivity, and overall faster analysis

for separating volatile organic compounds, such as those listed in U.S. EPA Method

8260B. The Rtx®-VMS stationary phase is a highly stable polymer that provides out-

standing analysis of volatile compounds, in combination with sensitive ion traps and

Agilent 5973 mass spectrometers. 0.18 and 0.25 mm ID columns allow sample splitting

at the injection port, eliminating the added expense and maintenance of a jet separator.

A 0.45 mm or 0.53 mm ID column can be directly connected to the purge-and-trap

transfer line in a system equipped with a jet separator.

Volatile Organics Analysis

GC_EV00428

Column

Rtx®-VMS, 20 m, 0.18 mm ID, 1.00 µm (cat.# 49914)

Sample

Diluent:

Water

Conc.:

10 ppb in 5 mL RO water (unless noted); ketones 2.5X

Injection

Purge and trap split (split ratio 40:1)

Liner:

1 mm split (cat.# 20973)

Inj. Temp.:

220 °C

Purge and Trap

Instrument:

Tekmar LCS 3100

Trap Type:

Vocarb® 3000

Purge:

11 min @ ambient, flow 40 mL/min

Dry Purge:

1 min, flow 40 mL/min

Desorb Preheat

Temp.:

245 °C

Desorb:

2 min @ 250 °C, flow 40 mL/min

Bake:

8 min @ 260 °C

Interface

Connection:

Injection port

Transfer Line

Tubing:

Silcosteel® treated 0.53 mm ID tubing (cat.# 20595)

Transfer Line

Temp.:

120 °C

Oven

Oven Temp.:

50 °C (hold 4 min) to 100 °C at 18 °C/min (hold 0 min) to 230 °C at

40 °C/min (hold 3 min)

Carrier Gas

He, constant flow

Flow Rate:

1.0 mL/min

Detector

MS

Mode:

Scan

Transfer Line

Temp.:

280 °C

Analyzer Type:

Quadrupole

Tune Type:

BFB

Ionization Mode: EI

Scan Range:

35-300 amu

Instrument

HP6890 GC & 5973 MSD

Notes

For proper flows, adjust retention time of dichlorodifluoromethane

to a retention time of 1.03 min @ 50 °C

¹A.L. Hilling and G. Smith, Environmental Testing & Analysis, 10(3), 15-19, 2001.

restek innovation!

• First choice for use with dual purge & traps

1

• EPA recommended surrogate used.

For peak listing,

visit

www.restek.com

and enter GC_EV00428 in the search.

30-Meter

60-Meter

75-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

cat.#

price

0.25 mm 1.40 µm -40 to 240/260 °C 19915

19916

0.32 mm 1.80 µm -40 to 240/260 °C 19919

19920

0.45 mm 2.55 µm -40 to 240/260 °C 19908

19909

0.53 mm 3.00 µm -40 to 240/260 °C 19985

19988

19974

20-Meter

40-Meter

ID

df

temp. limits

cat.#

price

cat.#

price

0.18 mm 1.00 µm -40 to 240/260 °C 49914

49915

Whole Air Canister Sampling

and Preconcentration

GC-MS Analysis

for pptv Levels of Trimethylsilanol

in Semiconduct

or

Cleanroom Air

lit. cat.#

EVAN1788-UNV

free literature

Download your free copies from

www.restek.com

Analysis of Trace Oxygenates

in Petroleum-Contaminated

Wastewater,

Using Purge-and-Trap

GC-MS (U.S. EPA Methods

5030B & 8260)

lit. cat.#

EVAN1449-UNV