Table of Contents Table of Contents
Previous Page  185 / 628 Next Page
Information
Show Menu
Previous Page 185 / 628 Next Page
Page Background www.restek.com

183

Paraquat and Diquat on Ultra Quat

LC COLUMNS

Application-Specific Columns

Chromatographic Properties

A retentive, high-purity, base-deactivated, reversed-phase packing. Ideal for the analysis of

paraquat and diquat or other quaternary amines.

Column Characteristics:

particle size:

3 µm or 5 µm, spherical

pore size:

100 Å

pH range:

2.5 to 8

temperature limit: 80 °C

Ultra Quat Columns

Ultra Quat Guard Cartridges

LC_EV0383

Column

Ultra Quat (cat.# 9181352)

Dimensions:

50 mm x 2.1 mm ID

Particle Size:

3 µm

Pore Size:

100 Å

Temp.:

ambient

Sample

Diluent:

DI Water

Conc.:

5 µg/mL each component

Inj. Vol.:

10 µL

Mobile Phase

10 mM heptafluorobutyric acid:acetonitrile (95:5)

Flow:

0.3 mL/min

Detector

Applied Biosystems/MDS Sciex LC-MS/MS

Model #:

API 3200™MS/MS system

Ion Source:

Electrospray

Ion Mode:

ESI+

Ion Spray Voltage:

5.5 kV

Curtain Gas:

15 psi (103.4 kPa)

Gas 1:

70 psi (482.6 kPa)

Gas 2:

60 psi (413.7 kPa)

Source Temp.:

600 °C

Mode:

MRM

Dwell Time:

200 ms

Instrument

Applied Biosystems/MDS Sciex LC-MS/MS System

Notes

Collision exit potential: 3V

Q1/Q3: unit resolution

Acknowledgement

Data courtesy of Houssain El Aribi, Ph.D., LC/MS Product and Application Specialist,

MDS SCIEX, 71 Four Valley Drive, Concord, Ontario, Canada, L4K 4V8

Peaks Precursor Ion Fragment Ion DP Collision Energy

(amu)

(amu)

(V)

(eV)

1. Diquat

183+

157+

30

30

2. Paraquat

93 (2+)

171+

20

20

Fast, sensitive

LC-MS/MS analysis!

Guard Cartridges

3-pk.

(10 x 2.1 mm)

3-pk.

(10 x 4.0 mm)

price

Ultra Quat Guard Cartridge

918150212

918150210

2.1 mm ID

4.6 mm ID

Length

cat.#

price

cat.#

price

3 µm Column

50 mm

9181352

5 µm Column

150 mm