![Show Menu](styles/mobile-menu.png)
![Page Background](./../common/page-substrates/page0183.jpg)
181
Organic Acids Standard on Allure® Organic Acids
LC COLUMNS
Application-Specific Columns
Column Characteristics:
particle size:
5 µm, spherical
pore size:
60 Å
end-cap:
no
pH range:
2.5 to 8
temperature limit: 80 °C
0
5
1
2
3
4
5
10
15
20
25
30
35
40 Time (min)
LC_0238
Column
Allure® Organic Acids (cat.# 9165585)
Dimensions:
300 mm x 4.6 mm ID
Particle Size:
5 µm
Pore Size:
60 Å
Temp.:
ambient
Sample
standard solution
Diluent:
water
Inj. Vol.:
10.0 µL
Mobile Phase
100 mM phosphate buffer, pH 2.5
Flow:
0.5 mL/min
Detector
UV/Vis @ 226 nm
Peaks
Conc.
(mg/mL)
1. Tartaric acid
1.0
2. Quinic acid
1.0
3. Malic acid
1.0
4. Citric acid
1.0
5. Fumaric acid
0.005
Chromatographic Properties
Allure® Organic Acids columns provide enhanced retention and selectivity for polar organic
acids, allowing the separation to be performed on a single 30 cm column. An Allure® Organic
Acids column effectively resolves key organic acids such as tartaric and quinic acids using
the chromatographic conditions specified in AOAC method 986.13. Retention is stable and
reproducible, even with the 100% aqueous mobile phase specified in the AOAC method.
Allure® Organic Acids Columns
Allure® Organic Acids Guard Cartridges
Guard Cartridges
3-pk.
(10 x 2.1 mm)
3-pk.
(10 x 4.0 mm)
price
Allure Organic Acids Guard Cartridge
916550212
916550210
3.0 mm ID
4.6 mm ID
Length
cat.#
price
cat.#
price
5 µm Column
150 mm
916556E
9165565
250 mm
—
9165575
300 mm
—